FDA Adverse Event
Malfunction
Summary report: N
MIDMARK DIGITAL EKG
MDR report key: 1693410
·
Received May 13, 2010
Report
- Report Number
- MW5015965
- Event Type
- Malfunction
- Date Received
- May 13, 2010
- Date of Event
- April 22, 2010
- Report Date
- May 13, 2010
- Manufacturer
- MIDMARK
- Product Code
- DRT
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- DE, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
IN A PEDIATRIC SETTING, WHEN A DIGITAL ECG IS COLLECTED, IT CORRECTLY CALCULATES THE QTC WITH THE BAZETT FORMULA. IF THE CARDIOLOGIST ELECTS TO CORRECT THE QT INTERVAL BY DOUBLE CLICKING ON "QT", THEY ARE PRESENTED WITH A SCREEN "EDIT MEASUREMENT". IF THEY SELECT "OK", -EVEN WITHOUT CHANGING THE VALUE-, THE HODGES FORMULA IS AUTOMATICALLY ACTIVATED RESULTING IN AN UNINTENTIONAL CHANGE IN THE QTC RECORDED.