FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 1693328 · Received May 18, 2010

Report

Report Number
2050012-2010-00273
Event Type
Malfunction
Date Received
May 18, 2010
Date of Event
May 4, 2010
Report Date
May 18, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
CHH
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE. FSE REPLACED SEVERAL HARDWARE COMPONENTS AND AS OF DATE, THERE HAVE BEEN NO FURTHER ISSUES REPORTED. ALTHOUGH HARDWARE FAILURES MAY HAVE CONTRIBUTED TO THIS EVENT, A CLEAR ROOT CAUSE FOR THIS EVENT CANNOT BE DETERMINED TO DATE.

Description of Event or Problem · 1

CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING INTERMITTENT ERRONEOUSLY LOW CHOLESTEROL (CHOL), TRIGLYCERIDES (TG), HDL CHOLESTEROL (HDLD) AND GLUCOSE (GLU) GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. INTERMITTENT LOW LIPID RESULTS WERE GENERATED AND REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN. SAMPLES WERE REPEATED ON A DIFFERENT INSTRUMENT AND AMENDED REPORTS WERE ISSUED. THERE WAS NO IMPACT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER CHH BECKMAN COULTER INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1