UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-00273
- Event Type
- Malfunction
- Date Received
- May 18, 2010
- Date of Event
- May 4, 2010
- Report Date
- May 18, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- CHH
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE. FSE REPLACED SEVERAL HARDWARE COMPONENTS AND AS OF DATE, THERE HAVE BEEN NO FURTHER ISSUES REPORTED. ALTHOUGH HARDWARE FAILURES MAY HAVE CONTRIBUTED TO THIS EVENT, A CLEAR ROOT CAUSE FOR THIS EVENT CANNOT BE DETERMINED TO DATE.
CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING INTERMITTENT ERRONEOUSLY LOW CHOLESTEROL (CHOL), TRIGLYCERIDES (TG), HDL CHOLESTEROL (HDLD) AND GLUCOSE (GLU) GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. INTERMITTENT LOW LIPID RESULTS WERE GENERATED AND REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN. SAMPLES WERE REPEATED ON A DIFFERENT INSTRUMENT AND AMENDED REPORTS WERE ISSUED. THERE WAS NO IMPACT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | CHH | BECKMAN COULTER INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |