FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1712K

MDR report key: 16930976 · Received May 15, 2023

Report

Report Number
2032227-2023-206964
Event Type
Malfunction
Date Received
May 15, 2023
Date of Event
April 30, 2023
Report Date
August 11, 2025
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
000000763000317157
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE NGP 640G INSULIN PUMP WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE NGP INSULIN PUMP, WHICH IS MARKETED IN THE UNITED STATES. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF-TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. THE CREST FIRMWARE AND THUS SOFTWARE WERE UNSUCCESSFUL DUE TO PUMP STUCK IN CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO BYPASS OPEN BOOK. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON PCBA 1, PCBA 2 AND FORCE SENSOR. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CRACKED KEYPAD OVERLAY, SCRATCHED CASE, BATTERY TUBE THREADS - CRACKED, CRACKED CASE-CORNER OF BELT CLIP RAILS AND CRACKED CASE (BATTERY TUBE). CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO MOISTURE DAMAGE ON ELECTRONIC ASSEMBLIES. UNABLE TO CONFIRM OTHER ALARMS/OTHER ANOMALIES DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO DOWNLOAD HISTORY FILES AND TRACES DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO CONFIRM HIGH BG'S. COSMETIC DAMAGE AT THE RETAINER RING WAS NOT CONFIRMED, HOWEVER, OTHER COSMETIC DAMAGE WAS CONFIRMED AT THE BATTERY COMPARTMENT. BLANK DISPLAY WAS NOT CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED INSULIN PUMP HAD PUMP ERROR 35. IT WAS REPORTED THAT THE CUSTOMER WENT SWIMMING, WHEN THEY CAME OUT OF THE POOL THEY NOTICED THAT PUMP WAS NOT TURNING ON AND ALSO FOUND A BIG CRACK ON THE BACK OF THE BATTERY COMPARTMENT. THE CUSTOMER ALSO STATED THAT THEY EXPERIENCED HYPERGLYCEMIA AS THE PUMP DID NOT WORK AT THAT MOMENT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED.  TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THERE WAS DAMAGE REPORTED ON THE PUMP'S RETAINER RING AND SILICONE CASE AS WELL. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP AND REVERT TO THE BACKUP PLAN AS PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1651168 640G INSULIN PUMP MMT-1712K PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1712K HG5SJFL 000000763000317157

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown