FDA Adverse Event Injury Summary report: N

FISHER & PAYKEL MYAIRVO 2 HUMIDIFIER

MDR report key: 16929604 · Received May 12, 2023

Report

Report Number
MW5117503
Event Type
Injury
Date Received
May 12, 2023
Date of Event
February 15, 2023
Report Date
May 10, 2023
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

PHONE CALL RECEIVED FROM INTAKE AT 0100 REGARDING PT (PATIENT) WITH BURNING CIRCUIT. TALKED W/ (WITH) MOM, (B)(6) WHO STATED THAT SHE SMELLED BURNING PLASTIC BUT HAD ASSUMED IT WAS NEIGHBOR, D/T (DUE TO) THEM HAVING WINDOW OPEN EARLIER IN NIGHT. (B)(6) WENT INTO PT (PATIENT) ROOM AND NOTICED THAT CIRCUIT PROXIMAL TO PT (PATIENT) WAS CRACKED. SHE ASKED HUSBAND TO WATCH PT (PATIENT) AS SHE WENT DOWN TO GET NEW CIRCUIT. WHEN SHE CAME BACK WITH NEW CIRCUIT, HUSBAND HAD STATED THAT CIRCUIT WAS MELTED TO THE BED SHEET. PARENTS UNHOOKED FROM AIRVO UNIT. MOM STATED THAT SHE UNHOOKED PT (PATIENT) FROM UNIT. ALSO, MOM STATED THAT THE CIRCUIT FELT VERY HOT. (B)(6) RT (RESPIRATORY THERAPIST) WENT OUT WITH NEW CIRCUIT AND NEW AIRVO TO EXCHANGE. UNIT EXCHANGED WITH MATCHING SETTINGS AS PER RX (PRESCRIPTION). ONCE AIRVO WARMED AND READY, HOOKED PT (PATIENT) BACK UP TO CIRCUIT. WATCHED PT (PATIENT) AND DISCUSSED EPISODE WITH MOM. SHE STATED THAT THE MACHINE NEVER ALARMED, SHE ONLY WENT INTO (B)(6) ROOM BECAUSE OF BURNT PLASTIC SMELL. VERIFIED THAT WATER WAS IN CANISTER, AND THAT NOTHING WAS COMPLETELY COVERING CIRCUIT. THE PT (PATIENT) SLEEPS IN FULL SIZE BED OVER TO THE LEFT SIDE, CIRCUIT WAS LAYING ON SHEET. MADE A MELTED COIL IMPRINT ON BEDDING. THERE WAS A LITTLE MELTED CIRCUIT ON A BLANKET NEARBY, BUT NOTHING WAS COVERED. EDUCATED MOM TO CALL MD (MEDICAL DOCTOR) IN MORNING TO EXPLAIN WHAT HAD HAPPENED. MOM SAID SHE WOULD FOLLOW UP AND NOTIFY ME OF ANY QUESTIONS/CONCERNS. PT O2 SATS (PATIENT OXYGEN SATURATION) AND HR WNL (HEART RATE WITHIN NORMAL LIMITS) BEFORE LEAVING. MOM CALLED IN AFTERNOON TO LET ME KNOW THAT PT (PATIENT) HAD DEVELOPED A COUGH AND WHEN SHE NOTIFIED HER PULMONOLOGIST, DR. (B)(6) THEY EXPRESSED THAT CONTINUE TO WATCH THE COUGH. EVENT ESCALATED TO (B)(6) MGR (MANAGER) AND INCIDENT REPORT DONE. EQUIPMENT BAGGED ALONG WITH CIRCUIT AND PUT UNDER EQUIPMENT AREA DESK. DID TAKE PICTURES AND EMAILED TO (B)(6) MGR (MANAGER). REFERENCE REPORT: MW5117504.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602940 FISHER & PAYKEL MYAIRVO 2 HUMIDIFIER HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) BTT FISHER & PAYKEL HEALTHCARE, LTD. PT100US 2100085684

Patients

Seq Age Sex Outcome Treatment
1 6 YR Female Hospitalization