FDA Adverse Event Other Summary report: N

STARBURST XLI ENHANCED SEM-FLEX, 25CM, FLEXIB

MDR report key: 1692856 · Received May 17, 2010

Report

Report Number
1056436-2010-00036
Event Type
Other
Date Received
May 17, 2010
Date of Event
May 7, 2010
Report Date
May 17, 2010
Manufacturer
ANGIODYNAMICS
Product Code
GEI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

NO INJURY. PT IS FINE. RFA OF THE LIVER. THEY WERE PLANNING TO DO A 5-6CM ABLATION. EXTENDED TINES OUT TO 5CM THEN SCAN. AFTER SCAN, RETRACT TINES BACK TO 2CM TO START ABLATION. WHEN RETRACTING BACK THEY FELT SOMETHING WAS NOT RIGHT. TOOK ANOTHER SCAN. THEY WERE ABLE TO TELL A TINE HAD BUSTED OFF. THEY ENDED UP DOING THE ABLATION. WHEN FINISHED IT WAS CONFIRMED THAT A TINE WAS BUSTED OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARBURST XLI ENHANCED SEM-FLEX, 25CM, FLEXIB RFA PROBE GEI ANGIODYNAMICS NA 507764

Patients

Seq Age Sex Outcome Treatment
1 UNK Other