FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 16921419 · Received May 12, 2023

Report

Report Number
1645337-2023-05674
Event Type
Injury
Date Received
May 12, 2023
Date of Event
April 19, 2023
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000518
PMA / PMN Number
P030053
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON JULY 11, 2023, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT'S IMPLANT WAS REPLACED WITH A 425CC MENTOR MEMORYGEL BREAST IMPLANT (CATALOG: 3504254BC LOT: 9840879 SN: (B)(6).

Additional Manufacturer Narrative · 0

ON SEPTEMBER 20, 2023, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE SUSPECT MEDICAL DEVICE HAS BEEN DISCARDED. THE CORRECT DATE OF BIRTH WAS ALSO PROVIDED AND IT HAS POPULATED. THE COMPLAINT DEVICE HAS BEEN DISCARDED. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND DEVICE MALFUNCTION CANNOT BE CONFIRMED.

Additional Manufacturer Narrative · 0

SECTION D 6B. EXPLANTATION DATE: (B)(6), 2023 SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 425CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE IV). AS A RESULT, THE PATIENT HAS BEEN SCHEDULED FOR BREAST IMPLANT REMOVAL SURGERY ON (B)(6), 2023.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1309882 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 3504254BC 7604050 00081317000518

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female Required Intervention