FDA Adverse Event Injury Summary report: N

SALTER LABS

MDR report key: 16919645 · Received May 12, 2023

Report

Report Number
3000219639-2023-00015
Event Type
Injury
Date Received
May 12, 2023
Date of Event
April 13, 2023
Report Date
April 12, 2023
Manufacturer
SALTER LABS
Product Code
CAT
UDI-DI
00607411915871
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THERE WAS AN INTERRUPTION OF THERAPY THAT COULD LEAD TO HYPOXIA. COMPLAINT HISTORY REVIEWED. THERE HAVE BEEN NO SIMILAR COMPLAINTS FOR 1600HF-25 CANNULA IN THE PREVIOUS 24 MONTHS. MORE INFORMATION SENT VIA TEXT MESSAGE AND PHONE CALL FROM CUSTOMER. CUSTOMER RECEIVES PRODUCT FROM MEDICAL SUPPLIER AS WELL AS PURCHASING FROM AMAZON. LOT# 354817 WAS BENT/KINKED IN PACKAGE. PURCHASED FROM AMAZON. SUNMED DOES NOT USE AMAZON AS A DISTRIBUTOR AND CANNOT CONTROL THE QUALITY OF PRODUCT DURING TRANSIT. MOST LIKELY ROOT CAUSE OF DEFECTS IS PRODUCT MISHANDLING DURING TRANSIT. RISK(RMA-20017A): R97: LOW OXYGEN DELIVERY - CANNULA MATERIAL DETERIORATES RESULTING IN LEAKS PRIOR TO FIRST USE - S=6 O=1 RPN=6. RPN < 25. R100: LOW OXYGEN DELIVERY - CANNULA TUBING DETERIORATES RESULTING IN TUBING COLLAPSE PRIOR TO FIRST USE - S=8 O=1 RPN=8. RPN < 25, THEREFORE RISK IS ACCEPTABLE. PRODUCT WAS PURCHASED THROUGH AN UNAPPROVED DISTRIBUTOR (AMAZON), ISSUE WILL NOT BE BROUGHT TO CARB.

Additional Manufacturer Narrative · 0

THERE WAS AN INTERRUPTION OF THERAPY THAT COULD LEAD TO HYPOXIA.

Description of Event or Problem · 0

KINKING IS PREVENTING FLOW OF OXYGEN. PATIENT CLAIMS THIS IS A LIFE THREATENING ISSUE AS THEY ARE NOT ABLE TO GET ADEQUATE OXYGEN.

Description of Event or Problem · 0

KINKING IS PREVENTING FLOW OF OXYGEN. PATIENT CLAIMS THIS IS A LIFE THREATENING ISSUE AS THEY ARE NOT ABLE TO GET ADEQUATE OXYGEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1662587 SALTER LABS ADULT NASAL CANNULA WITH 25', 3- CHANNEL TUBE CAT SALTER LABS 16SOFT-HF-25 UNKNOWN 00607411915871

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other