FDA Adverse Event
Death
Summary report: N
I-STAT CHEM8+ CARTRIDGE
MDR report key: 1691816
·
Received May 13, 2010
Report
- Report Number
- 2245578-2010-00028
- Event Type
- Death
- Date Received
- May 13, 2010
- Date of Event
- April 4, 2010
- Report Date
- May 12, 2010
- Manufacturer
- ABBOTT POINT OF CARE, ABBOTT POINT OF CARE CANADA LTD.
- Product Code
- MZV
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
ON 04/14/2010, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THE PT DIED ON (B) (6) 2010 FROM CHRONIC LYMPHOCYTIC LEUKEMIA (CLL). ON THE (B) (6) 2010, THE CUSTOMER REPORTED THE I-STAT CHEM8+ CARTRIDGE YIELDED A POTASSIUM RESULT 3.2 MMOL/L. SAME SAMPLE TESTED USING A LAB INSTRUMENT YIELDED AN POTASSIUM RESULT OF 6.7 MMOL/L, CUSTOMER STATED THE LAB SAMPLE WAS SLIGHTLY HEMOLYZED. WHITE BLOOD CELL COUNT OF THE PT WAS 300 000. THE CUSTOMER STATED, RESULTS DID NOT NECESSITATE MEDICAL OR SURGICAL INTERVENTION. ADDITIONAL INFO REQUESTED FROM THE CUSTOMER REGARDING THE DEATH OF THE PT AND TREATMENT, INFO NOT PROVIDED AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | MZV | ABBOTT POINT OF CARE, ABBOTT POINT OF CARE CANADA LTD. | E10050B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Death |