FDA Adverse Event Injury Summary report: N

ALLURA XPER FD

MDR report key: 16917725 · Received May 12, 2023

Report

Report Number
3003768277-2023-02831
Event Type
Injury
Date Received
May 12, 2023
Date of Event
May 4, 2023
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product Code
OWB
UDI-DI
00884838054189
PMA / PMN Number
K162859
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PHILIPS HAS INVESTIGATED THIS COMPLAINT. ACCORDING TO THE ADDITIONAL INFORMATION RECEIVED, THE STAFF MEMBER EXPERIENCED DIFFICULTIES WITH LONGITUDINAL AND LATERAL MOVEMENTS OF THE SYSTEM'S TABLE DURING THE TREATMENT OF A LARGE PATIENT (> 150 KG). THE NEXT MORNING, THE STAFF MEMBER EXPERIENCED PAIN AND NUMBNESS IN THE LEG, UNDERWENT AN MRI, AND WAS DIAGNOSED WITH A SLIGHT BULGING L4/5 DISC, AND SMALL TEARS WERE IDENTIFIED. SHE RECEIVED TREATMENT FOR THE INJURY AND FOUR MONTHS AFTER THE INCIDENT FULLY RETURNED TO WORK. A PHILIPS FIELD SERVICE ENGINEER (FSE) INSPECTED THE TABLE AND SYSTEM ONSITE AND DID NOT IDENTIFY ANY ISSUES. LOG FILE ANALYSIS DID NOT DISPLAY ANY DEVICE MALFUNCTION. PHILIPS CONFIRMED THAT THE SYSTEM WAS WORKING AS INTENDED AT THE TIME OF THE REPORTED EVENT AND THE CUSTOMER CONTINUED TO USE THE SYSTEM WITHOUT ANY FURTHER REPORTS OF SIMILAR ISSUES. STANDARD OF CARE DICTATES THAT HEALTH CARE PROFESSIONALS USE PROPER BODY MECHANICS AND ERGONOMICS DURING THE REPOSITIONING, TRANSFER AND TREATMENT OF PATIENTS. NO DETAILS OF THE STAFF MEMBER¿S BODY MECHANIC RELATED TO THE REPORTED MOVEMENT OF THE TABLE WERE RECEIVED BY PHILIPS.

Description of Event or Problem · 0

IT HAS BEEN REPORTED TO PHILIPS THAT THE PATIENT BED WAS TRICKY TO MOVE AND WHILE MOVING A LARGE PATIENT ONE MEMBER OF STAFF HURT THEIR BACK. SYSTEM WAS IN CLINICAL USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1185208 ALLURA XPER FD INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10 00884838054189

Patients

Seq Age Sex Outcome Treatment
1 40 YR Female Other