TRANSESOPHAGEAL ENDOSCOPIC TRANSDUCER
Report
- Report Number
- 3019216-2010-00003
- Event Type
- Other
- Date Received
- May 14, 2010
- Date of Event
- April 22, 2010
- Report Date
- April 29, 2010
- Manufacturer
- PHILIPS ULTRASOUND
- Product Code
- ITX
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THERE WAS NO REPORT OF SERIOUS INJURY WITH THIS EVENT. THE PROBE REMAINS AT THE HOSPITAL AND HAS NOT BEEN EVALUATED BY PHILIPS PERSONNEL. PHILIPS REQUESTED RECEIPT OF THE PROBE FOR EVALUATION. IT WAS CONFIRMED THE CUSTOMER IS USING A NON-APPROVED CHEMICAL (TRISTEL) TO STERILIZE THIS PROBE. PHILIPS USER DOCUMENTATION INSTRUCTS USERS TO USE A PROTECTIVE TRANSDUCER COVER DURING TEE EXAMS.
PHILIPS HEALTHCARE RECEIVED AN MHRA ADVERSE INCIDENT REPORT FROM USER FACILITY. (B)(6). THE DETAILS OF THE INCIDENT ARE DESCRIBED AS FOLLOWS: WHILE CLEANING THE PROBE AFTER USE IN A PATIENT AND PRIOR TO BEING STERILIZED, IT WAS NOTICED THAT THE O-RING USED TO SEAL THE AREA BETWEEN THE FLEXIBLE BENDING NECK AND THE SHAFT (INSERTION TUBE) WAS MISSING. IT WAS UNKNOWN WHEN AND WHERE THE O-RING BROKE OFF. NO APPARENT INJURY TO THE PATIENT WAS REPORTED. THE O-RING WAS OBSERVED MISSING AFTER USE IN A VERY SICK POST OP CARDIAC PATIENT AND THE PATIENT CURRENTLY REMAINS IN THE UNIT. THE ANAESTHETIST INVOLVED WITH THE PATIENT PERFORMED A MANUAL INSPECTION OF THE MOUTH AND OROPHARYNX AND PERFORMED A PERCUTANEOUS TRACHEOSTOMY UNDER DIRECT VISION WITH A BRONCHOSCOPE. AFTER INSERTION OF THE TRACHEOSTOMY, THE ANAESTHETIST VISUALIZED THE UPPER AIRWAY WITH THE BRONCHOSCOPE AND PERFORMED SUCTION OF THE AIRWAY SECRETIONS AND THE O-RING WAS NOT OBSERVED. THE PATIENT'S POOR CONDITION PREVENTED ANY FURTHER INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSESOPHAGEAL ENDOSCOPIC TRANSDUCER | X7-2T MTEE | ITX | PHILIPS ULTRASOUND | 453561193191 NB | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |