Narratives
PMA# OR 510K# : K050700. PMA# OR 510K#: K031049. ANTICIPATED PROCEDURAL COMPLICATION. THE SUBJECT DEVICE REMAINED IMPLANTED; THEREFORE, PHYSICAL ANALYSIS CANNOT BE PERFORMED. FROM THE INFORMATION AVAILABLE, IT WAS DETERMINED THAT THE DEVICE WAS USED IN ACCORDANCE WITH THE DIRECTION FOR USE, AND THAT THE USE OF THE COILS LIKELY CONTRIBUTED TO THE REPORTED PATIENT COMPLICATIONS. HOWEVER, THE DIRECTION FOR USE (DFU) STATES THAT MULTIPLE EMBOLIZATION PROCEDURES MAY BE REQUIRED TO ACHIEVE THE DESIRED OCCLUSION OF SOME ANEURYSMS OR VESSELS. THEREFORE A ROOT CAUSE OF ANTICIPATED PROCEDURAL COMPLICATION HAS BEEN ASSIGNED TO THIS EVENT.
THE PATIENT UNDERWENT A STENT ASSISTED COIL EMBOLIZATION OF A LEFT OPHTHALMIC ARTERY ANEURYSM THAT RESULTED IN THE PARTIAL OCCLUSION OF THE ANEURYSM. APPROXIMATELY 90 DAYS POST EMBOLIZATION, ANGIOGRAPHY DEMONSTRATED RECANALIZATION OF THE ANEURYSM SAC WITH COIL COMPACTION. SUCCESSFUL REINTERVENTION WAS PERFORMED 30 DAYS POST INDEX PROCEDURE. THE PATIENT WAS DISCHARGED SEVEN DAYS POST REINTERVENTION WITH GOOD RECOVERY AND IN STABLE CONDITION. NO ADDITIONAL INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MATRIX DETACHABLE COILS | DEVICE, NEUROVASCULAR EMBOLIZATION | HCG | BOSTON SCIENTIFIC - CORK | UNK536 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Required Intervention | (2) (B) (4) - MATRIX2 STANDARD| (3) (B) (4) - MATRIX2 STANDARD 3D| (3) (B) (4) - MATRIX2 STANDARD| 5) GDC 18 360 STANDARD| (1) (B) (4) - MATRIX2 STANDARD| (1) (B) (4) - MATRIX2 STANDARD| (1) (B) (4) -ATRIX2 SOFT 2D| (1) (B) (4) - NEUROFORM 3| (1) (CORDIS/J&J) ENTERPRISE STENT| (1) (B) (4) - MATRIX2 STANDARD| (1) (B) (4) - MATRIX2 ULTRASOFT| (4) (B) (4) -MATRIX2 STANDARD| (2) (B) (4) - MATRIX2 STANDARD| (2) (B) (4) -MATRIX2 SOFT 360| (BOSTON SCIENTIFIC) EXCELSIOR SL10 MICROCATHETER |