FDA Adverse Event Injury Summary report: N

MATRIX DETACHABLE COILS

MDR report key: 1691462 · Received May 17, 2010

Report

Report Number
2939204-2010-00121
Event Type
Injury
Date Received
May 17, 2010
Date of Event
February 24, 2005
Report Date
February 24, 2006
Manufacturer
BOSTON SCIENTIFIC - FREMONT
Product Code
HCG
PMA / PMN Number
UNKNOWN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B) (4) COIL PROTRUSION. PMA# OR 510K#: K012985 AND K031168. ANTICIPATED PROCEDURAL COMPLICATION. THE SUBJECT DEVICE REMAINED IMPLANTED; THEREFORE, PHYSICAL ANALYSIS CANNOT BE PERFORMED. FROM THE INFORMATION AVAILABLE, THERE IS NO INDICATION THAT THE REPORTED EVENT IS RELATED TO PRODUCT SPECIFICATION NONCONFORMANCE OR MISUSE. HOWEVER, COIL PROTRUSION AND NEUROLOGICAL SYMPTOMS ARE KNOWN AND ANTICIPATED COMPLICATIONS ASSOCIATED WITH COIL EMBOLIZATION PROCEDURES AND ARE LISTED AS SUCH IN THE DIRECTION FOR USE (DFU). THEREFORE, A ROOT CAUSE OF ANTICIPATED PROCEDURAL COMPLICATION HAS BEEN ASSIGNED TO THIS EVENT.

Description of Event or Problem · 1

DURING THE INDEX PROCEDURE, A TOTAL OF SEVEN COILS WERE DEPLOYED INTO THE LEFT ANTERIOR CEREBRAL ARTERY ANEURYSM. EMBOLISM AND COIL PROTRUSION WERE REPORTED AS TECHNICAL COMPLICATIONS ASSOCIATED WITH THE PROCEDURE. IMMEDIATELY AFTER TREATMENT THE SUBJECT WAS IMPROVED. IT WAS REPORTED THAT THE PATIENT UNDERWENT SUCCESSFUL REINTERVENTION EIGHT MONTHS (DATE UNKNOWN) POST INDEX PROCEDURE. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MATRIX DETACHABLE COILS DEVICE, NEUROVASCULAR EMBOLIZATION HCG BOSTON SCIENTIFIC - FREMONT UNK535

Patients

Seq Age Sex Outcome Treatment
1 53 YR Other| R (B) (4) - MATRIX STANDARD 3D| (B) (4) - MATRIX ULTRASOFT| (B) (4) - MATRIX ULTRASOFT| (B) (4) - MATRIX ULTRASOFT| (B) (4) - MATRIX FIRM 3D| EXCELSIOR 1018 MICROCATHETER (BOSTON SCIENTIFIC)| (B) (4) - MATRIX STANDARD-SR 2D| (B) (4) - MATRIX ULTRASOFT