FDA Adverse Event Malfunction Summary report: N

SPECIALITE

MDR report key: 1691259 · Received May 12, 2010

Report

Report Number
1824206-2010-07150
Event Type
Malfunction
Date Received
May 12, 2010
Date of Event
April 13, 2010
Report Date
April 13, 2010
Manufacturer
HILL-ROM INC.
Product Code
KZF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN REMOVED THE LEFT END CAP AND FOUND THAT THE CLEAR INSULATION AND WIRING HARNESS POWERING THE UPPER LIGHT WAS BURNT IN HALF. CUSTOMER DESCRIPTION: ON SUNDAY, (B) (6) 2010, AT ABOUT 11:30AM, THE LIGHT ABOVE THE BED IN PATIENT ROOM (B) (6), WAS OBSERVED SMOKING. THE NURSE IMMEDIATELY TURNED OFF THE LIGHT BY USING THE PULLCHAIN. THE PATIENT IN THE ROOM WAS MOVED TO ANOTHER ROOM. NO INJURIES. ON MONDAY, (B) (6) 2010, MAINTENANCE WAS NOTIFIED AND POWER TO THE LIGHT WAS SHUT OFF. ROOM IS VACANT AT THIS TIME. ADVISED MAINTENANCE TO INSPECT ALL WIRING HARNESSES. THE CUSTOMER SERVICE BULLETIN DATED (B) (6) 1988 WAS PROVIDED TO THE CUSTOMER. ACCOUNT STATED THAT HE HAS REPLACED THE WIRING HARNESS IN QUESTION AND THE LIGHT IS FUNCTIONING PROPERLY. NO FURTHER ISSUES. ACCOUNT IS REPLACING ALL OTHER HARNESSES AT THE FACILITY AS A PRECAUTION AS ROOMS BECOME AVAILABLE.

Description of Event or Problem · 1

ACCOUNT STATED THAT THE WIRES WHERE P640 LIGHT FLIPS OVER SMOKED AND BURNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECIALITE EXAM LIGHT KZF HILL-ROM INC. P645-00000618

Patients

Seq Age Sex Outcome Treatment
1 UNK