FDA Adverse Event Malfunction Summary report: N

RAYONE EMV

MDR report key: 16910175 · Received May 11, 2023

Report

Report Number
3012304651-2023-00064
Event Type
Malfunction
Date Received
May 11, 2023
Report Date
May 11, 2023
Manufacturer
RAYNER INTRAOCULAR LENSES LIMITED
Product Code
HQL
PMA / PMN Number
P060011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE REFERENCE (B)(4) HAS BEEN ALLOCATED TO THIS CASE BY RAYNER. THE EVENT DESCRIPTION PROVIDED STATES THAT ON AN UNKNOWN DATE FOLLOWING RAYONE EMV RAO200E IOL IMPLANTATION THE PATIENT PRESENTED WITH TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE AETIOLOGY OF TASS MAY BE MULTI-FACTORIAL WITH NUMEROUS POTENTIAL CAUSES INCLUDING BUT NOT LIMITED TO; BACTERIAL ENDOTOXINS OR PARTICULATE CONTAMINATION OF BALANCED SALT SOLUTIONS, INTRAOCULAR IRRIGATING SOLUTIONS WITH ABNORMAL PH, OSMOLARITY OR IONIC COMPOSITION, DENATURED OPHTHALMIC VISCOSURGICAL DEVICES (OVD), INTRAOCULAR MEDICATIONS (ANTIBIOTICS IN THE IRRIGATION SOLUTIONS OR INTRACAMERAL ANTIBIOTICS), TOPICAL OINTMENTS, INADEQUATE STERILIZATION OF SURGICAL INSTRUMENTS AND TUBING, INADEQUATE FLUSHING OF INSTRUMENTS BETWEEN CASES RESULTING IN BUILD-UP OF OPHTHALMIC VISCOSURGICAL DEVICES (OVD), PRESERVATIVE AND METALLIC PRECIPITATE. EVERY BATCH OF IOLS MANUFACTURED BY RAYNER ARE TESTED FOR ENDOTOXIN AND BIOBURDEN. RAYNER'S ENDOTOXIN ACCEPTABLE TOLERANCES ARE DERIVED FROM BS EN ISO 11979-08 (2017) OPHTHALMIC IMPLANTS - INTRAOCULAR LENSES PART 8 FUNDAMENTAL REQUIREMENTS. WE HAVE SET OUR OWN LIMITS FOR BIOBURDEN AS THERE IS NO STANDARD THAT PRESCRIBES ACCEPTABLE LIMITS. A RAYNER MICROBIOLOGICAL CAUSE FOR THE ONSET OF POST-OPERATIVE TASS IS THEREFORE VERY UNLIKELY. THERE IS INSUFFICIENT INFORMATION AVAILABLE CURRENTLY. RAYNER IS FOLLOWING UP VIA ITS AFFILIATE COMPANY TO OBTAIN ADDITIONAL INFORMATION TO FACILITATE FURTHER INVESTIGATION OF THE EVENT.

Description of Event or Problem · 0

ON 3RD MAY 2023, RAYNER RECEIVED NOTIFICATION FROM ITS US AFFILIATE COMPANY OF AN EVENT THAT OCCURRED FOLLOWING IMPLANTATION OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT POST-OPERATIVELY THE PATIENT PRESENTED WITH TOXIC ANTERIOR SEGMENT SYNDROME (TASS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1405726 RAYONE EMV RAYONE EMV HQL RAYNER INTRAOCULAR LENSES LIMITED RAO200E NOT AVAILABLE

Patients

Seq Age Sex Outcome Treatment
1 Unknown