FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1690995 · Received May 6, 2010

Report

Report Number
1824206-2010-06919
Event Type
Malfunction
Date Received
May 6, 2010
Date of Event
April 22, 2010
Report Date
April 22, 2010
Manufacturer
HILL-ROM INC
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

TECHNICIAN FOUND BED TAGGED "GCI INOPERABLE." HE FOUND POWER CONTROL BOARD TO BE INOPERATIVE. THE BED HAD ALL FUNCTIONS EXCEPT GCI. HE REPLACED PCB, BUT THIS DID NOT RESOLVE THE PROBLEM . HE REPLACED LEFT UCM CABLE AND BOARD AND THIS CORRECTED PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC P1900AF1619

Patients

Seq Age Sex Outcome Treatment
1