FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1690995
·
Received May 6, 2010
Report
- Report Number
- 1824206-2010-06919
- Event Type
- Malfunction
- Date Received
- May 6, 2010
- Date of Event
- April 22, 2010
- Report Date
- April 22, 2010
- Manufacturer
- HILL-ROM INC
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
TECHNICIAN FOUND BED TAGGED "GCI INOPERABLE." HE FOUND POWER CONTROL BOARD TO BE INOPERATIVE. THE BED HAD ALL FUNCTIONS EXCEPT GCI. HE REPLACED PCB, BUT THIS DID NOT RESOLVE THE PROBLEM . HE REPLACED LEFT UCM CABLE AND BOARD AND THIS CORRECTED PROBLEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC | P1900AF1619 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |