FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL XTRA BREAST IMPLANT

MDR report key: 16904007 · Received May 10, 2023

Report

Report Number
1645337-2023-05549
Event Type
Injury
Date Received
May 10, 2023
Date of Event
March 3, 2023
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317025016
PMA / PMN Number
P030053
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

USPECT DEVICE RECEIVED ON JUNE 28, 2023 DEVICE EVALUATION COMPLETED ON JULY 02 , 2023: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. DURING THE VISUAL EVALUATION, NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SMOOTH MOD + PROF XTRA 270CC RETURNED DEVICE. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. IT SHOULD BE NOTED THAT AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. PER ADDITIONAL INFORMATION RECEIVED WITH THE DEVICE, PATIENT UNDERWENT BILATERAL REPLACEMENTS AS FOLLOW; L)SN: (B)(6) / CAT:SMPX270, R) SN:(B)(6) / CAT:SMPX270 ON (B)(6) 2023. THE PROCEDURE TYPE WAS BREAST AUGMENTATION REVISION, PATIENT IDENTIFIER UPDATED TO PYL, AND DATE OF BIRTH UPDATED TO (B)(6) 1984. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

DEVICE EVALUATION COMPLETED ON APRIL 04 , 2023: UPON VISUAL EVALUATION OF THE IMAGE PROVIDED IN THE COMPLAINT NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. IT SHOULD BE NOTED THAT AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAP CON III. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A FEMALE WHO UNDERWENT AN UNSPECIFIED BREAST AUGMENTATION WITH A MENTOR MEMORYGEL XTRA, 270CC SILICONE PROSTHESIS EXPERIENCED BILATERAL CAPSULAR CONTRACTURE GRADE III POST PROCEDURE. THROUGH SONOGRAPHY, CAPSULAR CONTRACTURE BAKER'S GRADE 3 WAS DIAGNOSED IN BOTH BREASTS. AS A RESULT, PATIENT UNDERWENT BILATERAL REPLACEMENTS WITH UNKNOWN PROSTHESIS ON AN UNKNOWN DATE. THIS REPORT RELATES TO THE LEFT SIDE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
227801 MENTOR MEMORYGEL XTRA BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS SMPX270 9698854 00081317025016

Patients

Seq Age Sex Outcome Treatment
1 38 YR Female Required Intervention