FDA Adverse Event Malfunction Summary report: N

LEMAITRE VALVULOTOME

MDR report key: 16898949 · Received May 9, 2023

Report

Report Number
1220948-2023-00085
Event Type
Malfunction
Date Received
May 9, 2023
Date of Event
March 31, 2023
Report Date
May 9, 2023
Manufacturer
LEMAITRE VASCULAR, INC.
Product Code
MGZ
UDI-DI
00840663106653
PMA / PMN Number
K140042
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

WE HAVE RECEIVED THE COMPLAINT DEVICE FOR EVALUATION AND CONFIRMED THE REPORTED INCIDENT. WHEN THE GREEN HANDLE WAS PULLED TO CLOSE THE BLADES INTO ITS RETAINER SLOT ONE OF THE BLADES WAS PROTRUDING OUT OF THE RETAINER SLOT AND DIDN'T SIT PROPERLY ON THE SLOT. NO DAMAGE TO THE BLADE WAS FOUND. WE OBSERVED THAT THE BLADE WAS INCLINED OUTWARDS, PREVENTING IT FROM PROPERLY INSERTING INTO ITS RETAINER SLOT. PREVIOUS INVESTIGATIONS INTO THIS ISSUE HAVE FOUND THE FAILURE WAS DETERMINED TO BE CENTERING HOOP TO WIRE WELDING ERROR ALONG WITH OPERATOR ADJUSTMENT ERROR. DURING THE WELDING PROCESS, THIS CENTERING HOOP WAS LIKELY NOT PROPERLY ALIGNED AGAINST ITS RETAINER SLOT. THE BLADE PROCESS INCLUDES MANUAL MANIPULATION OF EACH BLADE BY THE OPERATOR. OPERATOR ERRORS ARE POSSIBLE DUE TO THE MANUAL NATURE OF THE ADJUSTMENTS. WE HAVE CONDUCTED A LOT HISTORY REVIEW AND DID NOT FIND ANY ISSUES NOTED IN THE MANUFACTURING OR PACKAGING PROCESS THAT COULD BE RELATED TO THIS ISSUE. ALL QC TESTS PASSED THEIR REQUIREMENTS. FURTHER, WE HAVE NOT RECEIVED ANY OTHER COMPLAINTS OF A SIMILAR NATURE FOR DEVICES FROM THIS LOT. CAPA 2022-025 WAS OPENED TO FURTHER INVESTIGATE THIS ISSUE.

Description of Event or Problem · 0

THE LEMAITRE VALVULOTOME COULD NO LONGER BE CLOSED. IT WAS REPORTED METAL THREADS ARE VISIBLE ON THE HEAD. THE BLADES BECAME UNABLE TO CLOSE. NO INJURY REPORTED TO THE PATIENT. THE PATIENT IS DOING FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231663 LEMAITRE VALVULOTOME VALVULOTOME MGZ LEMAITRE VASCULAR, INC. 1009-00 ELVH2087V 00840663106653

Patients

Seq Age Sex Outcome Treatment
1 Unknown