FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 16896291 · Received May 9, 2023

Report

Report Number
3013756811-2023-63358
Event Type
Malfunction
Date Received
May 9, 2023
Date of Event
April 17, 2023
Report Date
May 9, 2023
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613779
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. DEVICE NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A CARTRIDGE ALARM 1 OCCURRED. REPORTEDLY, THE CUSTOMER RELOADED THE CARTRIDGE TO RESOLVE THE ISSUE. ADDITIONALLY, IT WAS REPORTED THAT AN OCCLUSION ALARM OCCURRED. SYSTEM CHECK WAS PERFORMED BY TANDEM TECHNICAL SUPPORT AND NO ISSUES WERE IDENTIFIED. CUSTOMER CHANGED THE PUMP SUPPLIES AND RESUMED INSULIN DELIVERY. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 230-270 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1955168 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 W0884035 00850006613779

Patients

Seq Age Sex Outcome Treatment
1 27 YR Male INSULIN: HUMALOGINFUSION SET: AUTOSOFT 90