FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II)

MDR report key: 16887582 · Received May 8, 2023

Report

Report Number
1119779-2023-00534
Event Type
Malfunction
Date Received
May 8, 2023
Date of Event
April 19, 2023
Report Date
August 21, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSG
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 221261, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 3060999 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE BIOBURDEN TESTING. A SAMPLE OF PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. ALL BIOBURDEN TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER BD INTERNAL PROCEDURES. AFFECTED PRODUCT DOES NOT HAVE ANY STERILITY CLAIMS; THE PRODUCT IS TESTED FOR BIOBURDEN PRIOR TO RELEASE TO ENSURE THAT IT CONFORMS TO PRODUCT SPECIFICATIONS. HOWEVER, THIS DOES NOT ENSURE THAT THE END-USER WILL NOT RECEIVE A CONTAMINATED PLATE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 3060099. RETENTION SAMPLES FROM BATCH 3060099 WERE NOT AVAILABLE FOR INSPECTION. NO RETURN SAMPLES OR PHOTOS WERE RECEIVED FOR INVESTIGATION. THIS COMPLAINT CANNOT BE CONFIRMED. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION.

Additional Manufacturer Narrative · 0

D.3 COMMON DEVICE NAME: (B)(6). E.7: INITIAL REPORTER ADDR 1: (B)(6). H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: QUANTITY RECEIVED AND QUANTITY AFFECTED: 2 X 221261 - PLATE TRYPTICASE SOY AGAR 5% SB 100 EA AFFECTED. CUSTOMER REPORTED BIOLOGICAL CONTAMINATION ON LOT#: 3060099.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) THAT THERE WAS CONTAMINATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: QUANTITY RECEIVED AND QUANTITY AFFECTED: 2 X 221261 - PLATE TRYPTICASE SOY AGAR 5% SB 100 EA AFFECTED. CUSTOMER REPORTED BIOLOGICAL CONTAMINATION ON LOT#: 3060099.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
726078 BD BBL¿ TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) SEE H.10 JSG BECTON, DICKINSON & CO. (SPARKS) 3060099

Patients

Seq Age Sex Outcome Treatment
1 Unknown