FDA Adverse Event Injury Summary report: N

ULTRA FAST-FIX ASSEMBLY - CURVED

MDR report key: 16882565 · Received May 5, 2023

Report

Report Number
1219602-2023-00664
Event Type
Injury
Date Received
May 5, 2023
Date of Event
April 17, 2023
Report Date
January 10, 2024
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GAT
UDI-DI
03596010597205
PMA / PMN Number
K072322
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: H2: ADDITIONAL INFORMATION ON B5 AND H6. D4, LOT NO.: THE FOLLOWING ARE THE LOT NUMBERS REPORTED: 2066133, 2082806 AND 2083824. HOWEVER, IT IS UNKNOWN WHICH OF THE THREE HAS THE REPORTED MALFUNCTION. D4, EXP. DATE: EXPIRATION DATES FOR THE THREE LOTS REPORTED: 12-JAN-2026 (LOT 2066133 ), 24-NOV-2026 (LOT 2082806 ) AND 07-DEC-2026 (LOT 2083824). H4, MGF. DATE: MANUFACTURING DATES FOR THE THREE LOTS REPORTED: 12-JAN-2021 (LOT 2066133 ), 24-NOV-2021 (LOT 2082806 ) AND 07-DEC-2021 (LOT 2083824).

Additional Manufacturer Narrative · 0

H3, H6: PART OF THE DEVICE, USED IN TREATMENT, WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION REVEALED THE DEVICE WAS RETURNED IN ORIGINAL PACKAGING WITH BATCH NUMBER 2083824 ON THE LABEL. THE T1, T2, AND SUTURE WERE NOT RETURNED. THE VARIABLE DEPTH STRAW HAS BEEN TRIMMED. BIO DEBRIS IS PRESENT. A FUNCTIONAL EVALUATION COULD NOT BE PERFORMED AS THE IMPLANTS HAVE ALREADY BEEN DEPLOYED. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. THE INSTRUCTIONS FOR USE WAS REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THE ROOT CAUSE COULD NOT BE DETERMINED SINCE THE REPORTED MALFUNCTION COULD NOT BE DUPLICATED DURING THE INVESTIGATION. FACTOR THAT CAN CONTRIBUTE TO THE REPORTED EVENT INCLUDE A FAILURE TO FULLY ACTUATE THE DEVICE BEHIND THE MENISCUS DURING IMPLANTATION. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.

Additional Manufacturer Narrative · 0

H10: INTERNAL COMPLAINT REFERENCE: CASE (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A MENISCUS REPAIR, ONE (1) ULTRA FAST FIX FAILED, WHEN THE THREAD WAS PULLED, ONE OF THE T IMPLANTS CAME OUT. NO MATERIAL WAS LEFT INSIDE THE JOINT, SINCE THE OTHER T IMPLANT CAME OUT WHEN PULLING THE SUTURE AS WELL. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT STATUS IS GOOD.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A MENISCUS REPAIR, ONE (1) ULTRA FAST FIX FAILED. THE SURGICAL TECHNIQUE WAS PERFORMED AND BOTH T IMPLANTS REMAINED IN THE CAPSULE, BUT WHEN THE SPECIALIST PULLED THE THREAD TO CLOSE THE LOOP, THE T2 IMPLANT WAS RELEASED. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2212369 ULTRA FAST-FIX ASSEMBLY - CURVED SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE GAT SMITH & NEPHEW, INC. 72201491 UNKNOWN 03596010597205

Patients

Seq Age Sex Outcome Treatment
1 26 YR Male Other