BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2023-00161
- Event Type
- Malfunction
- Date Received
- May 5, 2023
- Date of Event
- April 14, 2023
- Report Date
- June 7, 2023
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL: YES. D10: RETURNED TO MANUFACTURER ON: 05-JUN-2023. H6: INVESTIGATION SUMMARY: ONE SAMPLE RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUAL EVALUATION, THE LUER THREAD IS BROKEN. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2211030, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. BASED ON THE TEAMS INVESTIGATION, POSSIBLE ROOT CAUSE IS ASSOCIATED WITH THE DAMAGE WAS LIKELY CAUSED AS A RESULT OF AN OVER SCREW OF THE LUER LOCK.
IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES THE THREAD OF THE DIFFUSER BROKE AND THE THREAD OF THE SYRINGE CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: INCIDENT DESCRIPTION: DURING THE INJECTION INTO A CHEMOTHERAPY DIFFUSER OF A 1ST VOLUME OF 50 ML OF GLUCOSE 5% USING A 3-PIECE SYRINGE OF 60 ML LUER LOCK: RAS. THEN DURING THE INJECTION OF THE 2ND VOLUME OF 5% GLUCOSE: THE THREAD OF THE DIFFUSER BROKE AND THE THREAD OF THE SYRINGE CRACKED.
IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES THE THREAD OF THE DIFFUSER BROKE AND THE THREAD OF THE SYRINGE CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: INCIDENT DESCRIPTION: DURING THE INJECTION INTO A CHEMOTHERAPY DIFFUSER OF A 1ST VOLUME OF 50 ML OF GLUCOSE 5% USING A 3-PIECE SYRINGE OF 60 ML LUER LOCK: RAS. THEN DURING THE INJECTION OF THE 2ND VOLUME OF 5% GLUCOSE: THE THREAD OF THE DIFFUSER BROKE AND THE THREAD OF THE SYRINGE CRACKED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1692065 | BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2211030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |