FDA Adverse Event Malfunction Summary report: N

VENTANA PD-L1 (SP142) ASSAY

MDR report key: 16875457 · Received May 5, 2023

Report

Report Number
2028492-2023-00028
Event Type
Malfunction
Date Received
May 5, 2023
Date of Event
April 20, 2023
Report Date
May 5, 2023
Manufacturer
VENTANA MEDICAL SYSTEMS INC
Product Code
PLS
PMA / PMN Number
P160002
Removal / Correction Number
Z-0907-2023 RES#91199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MATERIAL NUMBER IS NOT MARKETED OR DISTRIBUTED IN THE US. HOWEVER, ROCHE OBSERVED UNACCEPTABLE, LIGHT STAINING WITH SOME VENTANA PD-L1 (SP142) ON-MARKET LOTS, DURING INTERNAL COMPARISON STUDIES. LIGHT STAINING AFFECTS THE BORDERLINE OF POSITIVE VERSUS NEGATIVE TEST RESULTS. AN ON-GOING INVESTIGATION HAS DETERMINED THE ROOT CAUSE IS RELATED TO VARIABILITY IN THE SELECTION OF ANTIBODY CONCENTRATION IN RAW MATERIALS, AFFECTING SPECIFIC VENTANA PD-L1 (SP142) ASSAY LOTS MADE WITH THE IMPACTED RAW MATERIALS. A NOTIFICATION HAS BEEN SENT TO US CUSTOMERS INFORMING THEM OF THE ISSUE TO IMMEDIATELY DISCONTINUE THE USE OF AND DISCARD ANY REMAINING INVENTORY OF SPECIFIC IMPACTED LOTS AND INFORMING OF AN UPDATED DATE OF EXPIRATION FOR CERTAIN LOTS. INITIAL REPORTER FACILITY NAME TRUNCATED DUE TO CHARACTER LIMIT: ROYAL LONDON HOSPITAL CELLULAR PATHOLOGY 2ND FLOOR PATHOLOGY AND PHARMACY BLDG.

Description of Event or Problem · 0

A CUSTOMER FROM THE UNITED KINGDOM ALLEGED DISCREPANT RESULTS WITH THE VENTANA PD-L1 (SP142) ASSAY FOR 4 PATIENT SAMPLES. THE ALLEGED SAMPLES INITIALLY GENERATED NEGATIVE RESULTS WHICH WERE REPORTED TO MEDICAL PERSONNEL TREATING THE PATIENTS. THE SAMPLES WERE THEN RETESTED AND GENERATED POSITIVE RESULTS; THE RESULTS WERE SENT TO THE CLINICIANS. TO DATE, NO HARM OR INJURY IS ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1693279 VENTANA PD-L1 (SP142) ASSAY PROGRAMMED DEATH-LIGAND 1 (PD-L1) IVD, ANTIBODY PLS VENTANA MEDICAL SYSTEMS INC NA G16831,G24564

Patients

Seq Age Sex Outcome Treatment
1 Unknown