FDA Adverse Event Malfunction Summary report: N

ACTIVE LIFE

MDR report key: 16874264 · Received May 5, 2023

Report

Report Number
9618003-2023-00450
Event Type
Malfunction
Date Received
May 5, 2023
Report Date
April 10, 2023
Manufacturer
CONVATEC DOMINICAN REPUBLIC INC
Product Code
EXH
UDI-DI
00768455181339
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS EMDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER IMDRF CAUSE INVESTIGATION CODE, IMDRF INVESTIGATION FINDINGS, IMDRF CAUSE CONCLUSIONS. AN INVESTIGATION WAS COMPLETED FOR THIS COMPLAINT BY THE ASSIGNED MANUFACTURING SITE INVESTIGATION TEAM. THE COMPLAINT SUMMARY OF THIS INVESTIGATION REQUIRED REWORK. A CORRECTIVE ACTIONS/PREVENTIVE ACTIONS (CAPA) WAS RAISED AND AN UPDATED/CORRECTED INVESTIGATION SUMMARY HAS BEEN COMPLETED IN ACCORDANCE WITH ASSIGNED CAPA. A SUPPLEMENTAL MDR IS BEING SUBMITTED TO DOCUMENT THE COMPLETION OF INVESTIGATION REWORK AND FINAL INVESTIGATION; THE COMPLAINT RECORD WILL PROCEED TO CLOSURE. PHOTOGRAPH, VIDEO AND/OR PHYSICAL SAMPLE EVALUATION: NO PHOTOGRAPHS ASSOCIATED WITH THIS CASE WERE RECEIVED AND NO UNUSED RETURN SAMPLE WAS EXPECTED. BATCH RECORD REVISION RESULTS: LOT 2L02491 WAS MANUFACTURED ON 11/01/2022, IN CONVEX 1 PC LINE, WITH A TOTAL OF (B)(4) MARKET UNITS (MKUS). THE COMPLIANCE ENGINEER PERFORMED A BATCH RECORD REVIEW ON 02/OCT/2023, TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED AND NO ISSUES WERE FOUND; ALL THE COMPONENTS FOR ASSEMBLY WERE CORRECT PER BILL OF MATERIALS (BOM) AND ALL THE TOOLING INFORMATION DOCUMENTED WAS ALSO CORRECT, SYSTEM APPLICATION PRODUCT (SAP) MATERIAL ID 1000636 AND MANUFACTURING ORDER (B)(4). THE PRODUCTION PROCESS, IN-PROCESS CONTROL, TESTING RESULTS AND PACKAGING OF PRODUCTS WAS RUN ACCORDING TO THE PROCESS INSTRUCTION AND RECORDED IN BR31-146 VER. 7.0. REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WERE FOUND WITHIN THE DOCUMENTATION. HISTORICAL COMPLAINTS REVIEW: ON 02/OCT/2023, COMPLIANCE ENGINEER RAN A QUERY IN DATABASE IN ORDER TO VERIFY THE COMPLAINTS REPORTED FOR THE 2L02491 LOT FOR THE MALFUNCTION CODE ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (PRE-CUT ONLY)¿ AND AS RESULT, NO ADDITIONAL TYPE 2 COMPLAINTS WERE IDENTIFIED DURING THIS SEARCH AS PER WORK INSTRUCTION. HISTORICAL NONCONFORMANCE REVIEW: ON 02/OCT/2023, COMPLIANCE ENGINEER RAN A QUERY IN DATABASE LOOKING FOR ANY IN PROCESS NONCONFORMANCE / CORRECTIVE ACTION / PREVENTIVE ACTIONS (CAPA) (S) ASSOCIATED TO THE MALFUNCTION CODE ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (PRE-CUT ONLY)¿ FOR THE LOT NUMBER 2L02491 AND AS RESULT, NO NONCONFORMANCE / CORRECTIVE ACTION / PREVENTIVE ACTIONS (CAPA) (S) FOR THIS MALFUNCTION CODE WERE GENERATED DURING THE MANUFACTURING PROCESS OF THE REFERENCED LOT. CURRENT QUALITY CONTROLS: BASED ON THE PROCESS INSTRUCTION, THE FOLLOWING TESTS ARE PERFORMED IN THE MANUFACTURING LINES, IN ORDER TO IDENTIFY THIS FAILURE MODE IN OUR MANUFACTURING PROCESS: PD31-146 ¿DIMENSIONAL WAFER CTQ¿: FREQUENCY: 16 POUCHES PER TURN (2 POUCHES PER HOUR). SAMPLE QUANTITY: 16 POUCHES PER TURN (FIRST PIECE AT START) . ACCEPTANCE CRITERIA: ACCEPT = 0 / REJECT = 1 . DEFECT RATE ANALYSIS: THERE HAVE ONLY BEEN 1 DEFECTIVE PARTS CONFIRMED TO DATE FROM A LOT SIZE OF 2240 PRODUCTS. THIS REPRESENTS A DEFECT RATE OF ONLY (B)(4) , WHICH IS WELL WITHIN AN APPROPRIATE ACCEPTABLE QUALITY LEVEL (AQL) FOR THIS DEFECT WHICH SHOULD BE (B)(4) BASED ON OUR PD31-146 VER. 2.0 ¿CONVEX 1 PC SAMPLING PLAN¿. IN ADDITION, ALL OF THE IN-PROCESS TESTING ON THIS LOT DID NOT FIND A SINGLE DEFECTIVE POUCH, WHICH CONFIRMS THAT THE LOT IS UNLIKELY TO BREACH AN ACCEPTABLE QUALITY LEVEL (AQL) OF 2.5. THIS ISSUE CERTAINLY APPEARS TO BE AN ISOLATED INCIDENT, BUT MORE IMPORTANTLY TO DATE, IT IS WELL WITHIN OUR ACCEPTED ACCEPTABLE QUALITY LEVEL (AQL) LEVEL FOR THIS TYPE OF FAILURE MODE OR DEFECT. CONCLUSIONS: NO OBJECTIVE EVIDENCE WAS RECEIVED SUCH AS A PHOTO OR SAMPLE, FOR THIS REASON, WE CANNOT CONCLUDE THAT THE PRODUCT DOES NOT MEET SPECIFICATIONS. THE REVIEW OF THE BATCH RECORD FOR LOT 2L02491 SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS, ALL APPLICABLE MANUFACTURING AND QUALITY PROCESSES WERE FOLLOWED, AND NO DISCREPANCIES OR DEVIATIONS WERE RECORDED. NO CHANGES TO THE END-TO-END MANUFACTURING PROCESS OR COMPONENTS USED DURING ASSEMBLY OF THE BATCH WERE MADE. NO NONCONFORMITY HAS BEEN REGISTERED DURING THE MANUFACTURING PROCESS OF THE AFFECTED LOT FOR THE MALFUNCTION CODE ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (PRE-CUT ONLY)¿. NO ADDITIONAL COMPLAINTS WERE REPORTED FOR LOT AFFECTED RELATED TO THE MALFUNCTION CODE ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (PRE-CUT ONLY)¿. BASED ON THIS, NO NEGATIVE TREND WAS IDENTIFIED. BASED ON PRELIMINARY INVESTIGATION RESULTS, THERE IS NO OBJECTIVE EVIDENCE THAT OTHER PRODUCTS FROM THIS LOT ARE IMPACTED, AND THE ISSUE APPEARS TO BE ISOLATED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER: REPORTING SITE: 1049092. MANUFACTURING SITE: 9618003.

Additional Manufacturer Narrative · 0

DEVICE 6 OF 17. BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER, REPORTING SITE: 1049092. MANUFACTURING SITE: 9618003.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

THE END USER REPORTED THAT HE HAD FIVE MARKET UNITS OF COMPANY¿S KNOWN PRECUT OPENING POUCHES AND MOST OF THEM WERE OFF-CENTERED TO ONE DEGREE OR ANOTHER. HE STATED THAT APPROXIMATELY ONE-HALF TO TWO-THIRD WERE VERY OFF-CENTERED. HE ALSO MENTIONED THAT HE MIGHT BE ABLE TO USE JUST A HANDFUL FROM THESE FIVE BOXES. HE DID NOT WISH TO COUNT EVERY POUCH, BUT VERIFIED THAT HE HAD EXAMINED SOME IN EACH MARKET UNIT. FURTHERMORE, HE STATED THAT HE WILL SEND A PHOTO, HOWEVER AT THIS TIME, NO PHOTO RECEIVED. HE HAD NOT USED THE ONES THAT WERE OFF-CENTERED. HE ONLY USED THE ONE THAT HE FEELS LIKE WERE SAFE TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629289 ACTIVE LIFE BAG, URINARY, ILEOSTOMY EXH CONVATEC DOMINICAN REPUBLIC INC 175791 2L02491 00768455181339

Patients

Seq Age Sex Outcome Treatment
1 Unknown