FDA Adverse Event Malfunction Summary report: N

CAVITRON BOBCAT PRO 115V-G130(DNA)

MDR report key: 16872611 · Received May 4, 2023

Report

Report Number
2424472-2023-00021
Event Type
Malfunction
Date Received
May 4, 2023
Report Date
June 22, 2023
Manufacturer
DENTSPLY LLC
Product Code
ELC
PMA / PMN Number
K033705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE LACK OF WATER FLOW HAS CAUSED AN OVERHEATING INSERT. SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

EMN HP;CABLE,CONN/GUN CABLE KINK/PINCH/TWIST WATER LINE INSIDE THE HP CABLE TWISTED CAUSING RESTRICTING WATER FLOW TO THE HP. LEAKING WATER SOLENOID CAUSING CORROSION ON THE SCREWS OF THE HOUSING. HOUSING NEEDS TO BE BROKEN OPEN TO ACCESS INSIDE FOR EVALUATION. WILL REPLACE DAMAGED/WORN COMPONENTS AND RECALIBRATE UNIT TO FACTORY SPECS UPON ESTIMATE APPROVAL.

Description of Event or Problem · 0

WHILE USING A G130 CAVITRON BOBCAT PRO, THEY ALLEGE THAT THE HANDPIECE GETS HOT WHEN USING AND UNSURE IF WATER IS GETTING TO THE HANDPIECE. NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
432832 CAVITRON BOBCAT PRO 115V-G130(DNA) SCALER, ULTRASONIC ELC DENTSPLY LLC

Patients

Seq Age Sex Outcome Treatment
1 Unknown