FDA Adverse Event Malfunction Summary report: N

PUMP 1886 780G OUS BLE PUMP MG/DL

MDR report key: 16867926 · Received May 4, 2023

Report

Report Number
2032227-2023-203064
Event Type
Malfunction
Date Received
May 4, 2023
Date of Event
April 22, 2023
Report Date
August 18, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000521554
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF-TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON THE PCBA 1 / PCBA 2 / FORCE SENSOR, 40 PIN FLEXIBLE PRINTED CIRCUIT/LCD AND MOTOR. SUCCESSFULLY DOWNLOADED FIRMWARE / HISTORY FILES USING CREST AND THUS APPLICATION. PUMP ERROR 35 ALARM CONFIRMED IN HISTORY FILE OR TRACES (B)(6) 20223 16:59:49:000 PM DUE TO MOISTURE DAMAGE ON THE FORCE SENSOR. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION CRACKED BATTERY TUBE THREADS AND CRACKED CASE AT BATTERY TUBE SIDE. THE UNIT P-CAP / TEST RESERVOIR LOCKS IN PLACE PROPERLY. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO PUMP ERROR 35 ALARM. PUMP ERROR 35 ALARM CONFIRMED DUE TO MOISTURE DAMAGE. PUMP EXPOSED TO MOISTURE CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE NGP 780G INSULIN PUMP WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE NGP INSULIN PUMP, WHICH IS MARKETED IN THE UNITED STATES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED THE ERROR CODE OF A BROKEN FORCE SENSOR WAS DETECTED DURING SEATING OR REGULAR DELIVERY (PUMP ERROR 35). TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE INSULIN PUMP PERFORMED SAFETY CHECKS AND THE ERROR WAS FOUND. IT WAS FOUND THAT THE PUMP HAD OTHER CRITICAL PUMP ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE DEVICE AND REVERT TO THE BACKUP PLAN AS PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
859886 PUMP 1886 780G OUS BLE PUMP MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1886 HG62FRSZZ 000000763000521554

Patients

Seq Age Sex Outcome Treatment
1 Unknown