FDA Adverse Event Malfunction Summary report: N

ACHIEVE MAPPING CATHETER - 20 MM

MDR report key: 16860893 · Received May 3, 2023

Report

Report Number
9617601-2023-00096
Event Type
Malfunction
Date Received
May 3, 2023
Date of Event
February 15, 2023
Report Date
May 3, 2023
Manufacturer
MEDTRONIC MEXICO S. DE R.L. DE CV
Product Code
DRF
PMA / PMN Number
K102588
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE 990036-020 MAPPING CATHETER WITH LOT NUMBER 223228877 WAS RETURNED AND ANALYZED. VISUAL INSPECTION OF THE LOOP SEGMENT AREA SHOWED THE LOOP WAS INTACT WITH NO APPARENT ISSUES. NO DAMAGE WAS OBSERVED ALONG WITH THE TIP/LOOP SECTION OF THE MAPPING CATHETER. VISUAL INSPECTION OF THE PEBAX SEGMENT AREA SHOWED THE PEBAX TUBING WAS KINKED/TWISTED AT APPROXIMATELY 3 INCHES FROM LOOP DISTAL TIP. THE USER MAY HAVE EXPERIENCED INSERTION DIFFICULTIES DUE TO THIS ISSUE WHEN INTRODUCING THE MAPPING CATHETER INTO THE BALLOON CATHETER. VISUAL INSPECTION OF THE ELECTRODES SHOWED THEY WERE INTACT WITH NO APPARENT ISSUES. ALL ELECTRODES EXIST ON THE LOOP SECTION AND NO COSMETIC ISSUE OR ANOMALIES WERE IDENTIFIED. VISUAL INSPECTION OF THE SHAFT SEGMENT AREA SHOWED THE SHAFT WAS KINKED APPROXIMATELY 5.5 INCHES FROM THE LEMO CONNECTOR. VISUAL INSPECTION OF THE INTRODUCER SHOWED THE INTRODUCER WAS INTACT WITH NO APPARENT ISSUES. NO DAMAGE OR ANY OTHER ISSUE WAS OBSERVED ALONG WITH THE INTRODUCER. VISUAL INSPECTION OF THE RETURNED PACKAGE SHOWED THE PACKAGE WAS INTACT WITH NO APPARENT ISSUES. NO ANOMALIES WERE IDENTIFIED. THE FUNCTIONAL TEST WAS PERFORMED USING A MULTIMETER THE MAPPING CATHETER WAS CONNECTED TO THE TEST CABLE. THE CONTINUITY AND IMPEDANCE MEASUREMENT BETWEEN THE ELECTRODES AND THE OTHER SIDE OF THE CABLE SHOWED THE ELECTRODE'S CONTINUITY AND IMPEDANCE TO THE CABLE ARE NORMAL. IN CONCLUSION, THE MAPPING CATHETER FAILED THE RETURNED PRODUCT INSPECTION DUE TO A KINK/TWIST OBSERVED ON THE CATHETER SHAFT. THE REPORTED "ECG SIGNAL ISSUE (NO SIGNAL/NOISE/ARTIFACTS)" WAS NOT OBSERVED/REPRODUCED WITH THE RETURNED CATHETER. A RIBBED MATERIAL OBSERVED ON THE PEBAX TUBING. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE MAPPING CATHETER WAS EXPERIENCING SIGNAL ISSUE AND THE POTENTIAL COULD NOT BE DISPLAYED. THE MAPPING CATHETER WAS REPLACED TO RESOLVE THE ISSUE. THE CASE WAS COMPLETED WITH CRYO. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621272 ACHIEVE MAPPING CATHETER - 20 MM CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT DRF MEDTRONIC MEXICO S. DE R.L. DE CV 990063-020 223228877

Patients

Seq Age Sex Outcome Treatment
1 Unknown