FDA Adverse Event Injury Summary report: N

SMOOTH ROUND MOD. PLUS PROFILE

MDR report key: 16857861 · Received May 2, 2023

Report

Report Number
1645337-2023-05157
Event Type
Injury
Date Received
May 2, 2023
Date of Event
March 21, 2023
Manufacturer
MENTOR MEDICAL SYSTEM B.V.
Product Code
FTR
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE, MATERIAL RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON MAY 17, 2023, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. PER RETURNED DEVICE, MENTOR BECAME AWARE THAT IT'S LOT NUMBER WAS 6421693. THIS DEVICE WAS MANUFACTURE BY MENTOR MEDICAL SYSTEMS B.V, LOCATED IN LEIDEN, THE NETHERLANDS. IT IS A FOREIGN MANUFACTURER OF MEDICAL DEVICES THAT ARE NOT CLEARED OR APPROVED FOR SALE IN THE US. IT IS A SEPARATE LEGAL ENTITY FROM MENTOR TEXAS. PER 21 CFR 803.58 AND "MEDICAL DEVICE REPORTING FOR MANUFACTURERS" (FDA GUIDANCE DOCUMENT ISSUED ON NOVEMBER 8, 2016), WE ARE ¿[NOT] REQUIRED TO SUBMIT MDR REPORTS FOR EVENTS OCCURRING IN OTHER COUNTRIES FOR A DEVICE THAT IS MANUFACTURED IN A FOREIGN COUNTRY AND THAT IS NOT CLEARED OR APPROVED FOR MARKETING IN THE US. THEREFORE, THIS WILL BE THE LAST REPORT FOR THIS EVENT. NO FURTHER SUPPLEMENTAL REPORT WILL BE SUBMITTED FOR THIS COMPLAINT FILE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT BREAST AUGMENTATION WITH TWO 550CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE III) AND LEFT BREAST IMPLANT RUPTURE. AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2023. THE REPLACEMENT DEVICES WERE TWO: 650CC MENTOR MEMORYGEL BREAST IMPLANT (CATALOG: 3506504BC). THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
246508 SMOOTH ROUND MOD. PLUS PROFILE PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR MEDICAL SYSTEM B.V. 3505501BC 6421693

Patients

Seq Age Sex Outcome Treatment
1 49 YR Female Required Intervention