PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM
Report
- Report Number
- 2024168-2023-04609
- Event Type
- Injury
- Date Received
- May 2, 2023
- Date of Event
- April 13, 2023
- Report Date
- July 13, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE OTHER PERCLOSE PROSTYLE DEVICES REFERENCED IN B5 ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.
ANALYSIS AND SCANNING ELECTRON MICROSCOPY WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. ADDITIONALLY, RETURNED DEVICE ANALYSIS DEEMED THE POSTERIOR CUFF TABS WERE NOT BENT AS EXPECTED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE REPORTED NEEDLE TO CUFF MISS WAS CONCLUDED TO BE RELATED TO A POTENTIAL PRODUCT QUALITY ISSUE DUE TO THE POSTERIOR CUFF TABS NOT BEING BENT AS EXPECTED. THE ISSUE IS BEING ADDRESSED PER INTERNAL OPERATING PROCEDURES. ABBOTT VASCULAR WILL CONTINUE TO TREND THE PERFORMANCE OF THESE DEVICES. THE SUBSEQUENT TREATMENTS APPEAR TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE.D4; LOT NUMBER CHANGED FROM 2121441 TO 3010942.
IT WAS REPORTED THAT SUTURE PLACEMENT IN THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH FOUR PROSTYLE DEVICES USING THE PRE-CLOSE TECHNIQUE VIA AN 8F SHEATH HOLE PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) INTERVENTIONAL PROCEDURE. REPORTEDLY, WITH ALL 4 DEVICES, WHEN THE PLUNGER WAS PULLED BACK IN STEP 3, THE SUTURE WAS NOT PRESENT. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 14F AND THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 160827 | PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 3010942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |