DAVINCI XI
Report
- Report Number
- 2955842-2023-12332
- Event Type
- Malfunction
- Date Received
- May 2, 2023
- Date of Event
- March 16, 2023
- Report Date
- April 6, 2023
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110744
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
BASED ON THE CLAIM AGAINST THE PRODUCT BY THE CUSTOMER NOTING THE RIGHT YELLOW PEDAL WORKING INTERMITTENTLY, AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE WAS UNABLE TO REPLICATE THE REPORTED ISSUE; HOWEVER REPLACED THE FOOTREST TO RESOLVE THE REPORTED ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE FOOTREST WAS ANALYZED AND FAILURE ANALYSIS INVESTIGATION WAS ABLE TO REPLICATE/CONFIRM THE CUSTOMER REPORTED ISSUE. FAILURE ANALYSIS INSTALLED FOOTREST INTO PCA XI SYSTEM AND FOUND THE RIGHT YELLOW PEDAL IS NOT FUNCTIONAL PROPERLY WHEN PRESSED. THIS COMPLAINT IS REPORTABLE MALFUNCTION EVENT DUE TO THE FOLLOWING CONCLUSION: THE CUSTOMER ABORTED TO ANOTHER SYSTEM AFTER THE START OF THE PROCEDURE DUE TO INTERMITTENT FOOT PEDAL ISSUES. THE FSE REPLACED THE FOOTREST TO RESOLVE THE REPORTED ISSUE. FAILURE ANALYSIS WAS ABLE TO REPLICATE THE CUSTOMER REPORTED ISSUE. WHILE THERE WAS NO HARM OR INJURY TO THE PATIENT, SYSTEM UNAVAILABILITY AFTER THE START OF A SURGICAL PROCEDURE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF IT WERE TO RECUR AS IT MAY LEAD TO AN INJURY DUE TO THE PATIENT¿S INABILITY TO TOLERATE A PROCEDURE CHANGE.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LIVER RESECTION SURGICAL PROCEDURE, THE RIGHT YELLOW PEDAL WAS WORKING INTERMITTENTLY ON THE SURGEON SIDE CONSOLE (SSC) 1. INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) REVIEWED THE SYSTEM LOGS AND FOUND NO ERRORS. PRIOR TO CALLING, THE SURGEON SWITCHED TO THE SURGEON SIDE CONSOLE (SSC) 2. THE TSE SUGGESTED TO USE COMPRESSED AIR TO BLOW UNDER THE PEDAL TO POSSIBLY CLEAN OUT DUST AND DEBRIS FROM UNDER THE PEDAL. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER ON AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PROCEDURE WAS COMPLETED ROBOTICALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1533557 | DAVINCI XI | SURGEON SIDE CONSOLE, SMART PEDALS | NAY | INTUITIVE SURGICAL, INC | 380677-06 | N/A | 00886874110744 |