FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 16854447 · Received May 2, 2023

Report

Report Number
1645337-2023-05098
Event Type
Injury
Date Received
May 2, 2023
Date of Event
December 29, 2021
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000488
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE, MATERIAL RUPTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON MAY 3, 2023, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON MAY 10, 2023, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR THEN CONDUCTED VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE SMOOTH HPG, 350CC BREAST IMPLANT WAS FOUND TO BE RUPTURED AND RECEIVED IN THREE PARTS. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN AN AREA OF THE TEAR ON THE POSTERIOR ASPECT, MEASURING APPROXIMATELY LESS THAN 0.1 CM. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. THE CAUSE OF THE RUPTURE IN THE REMAINING AREA OF THE TEAR COULD NOT BE IDENTIFIED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. CAPA ACTIVITY WAS NOT REQUIRED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A 49-YEAR OLD CAUCASIAN FEMALE PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 350CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE IV ON THE RIGHT AND BAKER GRADE II ON THE LEFT). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2023. UPON REMOVAL, THE IMPLANTS WERE IDENTIFIED TO BE RUPTURED. THE REPLACEMENT DEVICES WERE: (RIGHT) 355CC MENTOR MEMORYGEL XTRA BREAST IMPLANT CATALOG: SHPX355, LOT: 9880785, SN:(B)(6), AND (LEFT) 355CC MENTOR MEMORYGEL XTRA BREAST IMPLANT CATALOG: SHPX355, LOT: 9880785, SN: (B)(6). THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1800250 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 3503504BC 6937206 00081317000488

Patients

Seq Age Sex Outcome Treatment
1 49 YR Female Required Intervention