FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II¿ PRN Y ADAPTER

MDR report key: 16850478 · Received May 1, 2023

Report

Report Number
3014704491-2023-00161
Event Type
Malfunction
Date Received
May 1, 2023
Date of Event
April 1, 2023
Report Date
May 8, 2023
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903830695
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2082668. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD INTIMA-II¿ PRN Y ADAPTER LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: THE NURSE ON DUTY FOUND THAT THE HOSE OF THE INDWELLING NEEDLE WAS LEAKING WHILE INFUSING THE PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD INTIMA-II¿ PRN Y ADAPTER LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: THE NURSE ON DUTY FOUND THAT THE HOSE OF THE INDWELLING NEEDLE WAS LEAKING WHILE INFUSING THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1711500 BD INTIMA-II¿ PRN Y ADAPTER INTRAVASCUALR CATHETER FOZ BD (SUZHOU) 2082668 00382903830695

Patients

Seq Age Sex Outcome Treatment
1 Unknown