FDA Adverse Event Malfunction Summary report: N

INFANT HEEL WARMER SQUEEZE 4X4IN

MDR report key: 16845267 · Received May 1, 2023

Report

Report Number
1423537-2023-00920
Event Type
Malfunction
Date Received
May 1, 2023
Date of Event
April 5, 2023
Report Date
June 27, 2023
Manufacturer
MO016 MO-MOBERLY
Product Code
MPO
UDI-DI
10192253061420
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CARDINAL HEALTH HAS RECEIVED AN INCREASE IN BURST/LEAKING COMPLAINTS REGARDING THE NOVALPLUS INFANT HEEL WARMER WITH TAPE. CUSTOMER COMPLAINTS HAVE NOTED RUPTURING AND BURSTING OF THE DEVICE WHEN ACTIVATION IS ATTEMPTED, MAKING THE DEVICE UNUSABLE AND IN MANY CASES CAUSING THE CONTENTS INSIDE THE POUCH TO LEAK AND CONTACT CLINICIANS, PATIENTS, AND OTHER INDIVIDUALS OR SURFACES IN PROXIMITY. SAMPLES RETURNED TO CARDINAL HEALTH FROM CUSTOMERS HAVE BEEN EVIDENCED TO HAVE AN INCOMPLETE TOP SEAL, RESULTING IN THE MALFUNCTION. THERE HAVE BEEN NO REPORTABLE ADVERSE EVENTS/INJURIES ASSOCIATED WITH V11460-010, LOT V2S056. CARDINAL HEALTH HAS HALTED DISTRIBUTION OF THE NOVALPLUS INFANT HEEL WARMER W/TAPE FOR LOT NUMBER V2S056 AND INITIATED A VOLUNTARY RECALL ON JUNE 16, 2023.

Additional Manufacturer Narrative · 0

WE RECEIVED TWO SAMPLES FOR INVESTIGATION THAT WERE NOT ALREADY ACTIVATED. ONE SAMPLE WAS DETERMINED TO HAVE AN INCOMPLETE TOP SEAL DUE TO MACHINE SETUP. THE ROOT CAUSE WAS DETERMINED TO BE A BURST PACK DUE TO AN INCOMPLETE TOP SEAL. THE MACHINE WAS SETUP CORRECTLY AND IS DESIGNED WITH A MECHANISM WHICH REDUCES THE LIKELIHOOD FOR THIS ISSUE. DEVICE HISTORY RECORD REVIEW WAS COMPLETED ON THE REPORTED LOT V2S056. THE LOT WAS FOUND TO HAVE BEEN MANUFACTURED AND RELEASED TO PREDETERMINED SPECIFICATIONS. NO ANOMALIES WERE FOUND DURING REVIEW OF THE RECORDS.

Description of Event or Problem · 0

CARDINAL HEALTH INFANT HEEL WARMER, BURST WHILE NURSE WAS ACTIVATING. SHE DID NOT SEEK ANY MEDICAL CARE AS SHE STATES THE CONTENTS OF THE HEEL WARMER LANDED MAINLY ON HER CHEST, DOWN HER UNIFORM AND ON HER PANTS. SHE WAS WEARING GLOVES AND CONTENTS DID NOT MAKE CONTACT WITH SKIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
737378 INFANT HEEL WARMER SQUEEZE 4X4IN INFANT HEEL WARMER (CHEMICAL HEAT PACK) MPO MO016 MO-MOBERLY 11460-010T V2S056 10192253061420

Patients

Seq Age Sex Outcome Treatment
1 Female