FDA Adverse Event Malfunction Summary report: N

RESONATE X4 CRT-D

MDR report key: 16843515 · Received May 1, 2023

Report

Report Number
2124215-2023-20744
Event Type
Malfunction
Date Received
May 1, 2023
Date of Event
March 3, 2023
Report Date
May 5, 2023
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
NIK
UDI-DI
00802526589539
PMA / PMN Number
P010012/S436
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IF PERTINENT INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CORRECTION TO H6: DEVICE CODE A070102/HIGH CAPTURE THRESHOLD WAS REMOVED.

Additional Manufacturer Narrative · 0

IF PERTINENT INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED FARFIELD OVERSENSING. IN ADDITION, THERE WAS AN ALERT TO CHECK THE RIGHT ATRIAL (RA) LEAD. REVIEW OF RA OUPUT REVEALED AUTO [email protected]; TECHNICAL SERVICES (TS) EXPLAINED THIS WAS LIKELY ATTRIBUTED TO FAILED RA AUTO THRESHOLD (RAAT) TESTS DUE TO THE OBSERVED FARFIELD OVESENSING. THIS CRT-D REMAINS IN SERVICE, AND DEVICE SENSITIVITY WAS ADJUSTED TO ADDRESS THE FARFIELD SIGNALS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED FARFIELD OVERSENSING. IN ADDITION, THERE WAS AN ALERT TO CHECK THE RIGHT ATRIAL (RA) LEAD. REVIEW OF RA OUPUT REVEALED AUTO [email protected]; TECHNICAL SERVICES (TS) EXPLAINED THIS WAS LIKELY ATTRIBUTED TO FAILED RA AUTO THRESHOLD (RAAT) TESTS DUE TO THE OBSERVED FARFIELD OVESENSING. THIS CRT-D REMAINS IN SERVICE, AND DEVICE SENSITIVITY WAS ADJUSTED TO ADDRESS THE FARFIELD SIGNALS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1362291 RESONATE X4 CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D NIK BOSTON SCIENTIFIC CORPORATION G447 277815 00802526589539

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male