BD DISCARDIT¿ II - 2-PIECE SYRINGE
Report
- Report Number
- 2243072-2023-00666
- Event Type
- Malfunction
- Date Received
- April 28, 2023
- Date of Event
- March 31, 2023
- Report Date
- May 3, 2023
- Manufacturer
- BECTON DICKINSON
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: PHOTOS WERE RECEIVED BY BD FOR EVALUATION. A QUALITY ENGINEER WAS ABLE TO REVIEW THE PHOTO FOR THE REPORTED COMPLAINTS OF MARKING DEFECT, WITHOUT SYRINGE FROM LOT NUMBER 2061254 REGARDING MATERIAL NUMBER 301866. THE PHOTOGRAPHS CONFIRMED THE REPORTED DEFECTS ON THE SYRINGE. THE INVESTIGATING TEAM HAS USED THE RETENTION SAMPLES OF LOT NUMBER 2061254 REGARDING MATERIAL NUMBER 301866 FOR INVESTIGATING THE REPORTED DEFECT. THE SAMPLES WERE VISUALLY TESTED, AND NO DEFECT WAS FOUND IN THE RETENTION SAMPLES. THE EXACT ROOT CAUSE CAN ONLY BE DETERMINED IF WE RECEIVE THE ORIGINAL SAMPLE. H3 OTHER TEXT : SEE H10.
THE MANUFACTURING LOCATION FOR THIS PRODUCT IS BAWAL, INDIA. THIS SITE IS NOT REGISTERED WITH THE FDA. THEREFORE, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AND THE FRANKLIN LAKES FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. E1: ENTER ADDRESS INFORMATION: (B)(6).
IT WAS REPORTED THAT 289 OF THE BD DISCARDIT¿ II - 2-PIECE SYRINGE EXPERIENCED SCALE MARKING ISSUES. 1 OF 2. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MARKING DEFECTS.
IT WAS REPORTED THAT 289 OF THE BD DISCARDIT¿ II - 2-PIECE SYRINGE EXPERIENCED SCALE MARKING ISSUES. 1 OF 2. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MARKING DEFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1046727 | BD DISCARDIT¿ II - 2-PIECE SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON | 2061254 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |