FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 16830873 · Received April 28, 2023

Report

Report Number
3013756811-2023-59451
Event Type
Malfunction
Date Received
April 28, 2023
Date of Event
April 8, 2023
Report Date
April 28, 2023
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613762
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN O-RING WAS ON THE PNEUMATIC TAP. THE CUSTOMER WAS ABLE TO REMOVE THE O-RING FROM THE PUMP AND SUCCESSFULLY LOADED A NEW CARTRIDGE TO RESOLVE THIS ISSUE. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 250-299 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430298 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00850006613762

Patients

Seq Age Sex Outcome Treatment
1 20 YR Female