FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM

MDR report key: 16830415 · Received April 28, 2023

Report

Report Number
0002023141-2023-01192
Event Type
Injury
Date Received
April 28, 2023
Date of Event
January 31, 2023
Report Date
April 28, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019522
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). PMA/510(K) NUMBER :K01102 AND K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMITIES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

DOCTOR REPORTED PERI-IMPLANTITIS WITH DISCOMFORT AND PAIN AT TOOTH LOCATION 13.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476187 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB13 1251488 00889024019522

Patients

Seq Age Sex Outcome Treatment
1 55 YR Male Required Intervention