FDA Adverse Event Malfunction Summary report: N

TRITON CANISTER

MDR report key: 16829166 · Received April 28, 2023

Report

Report Number
3011157718-2023-00047
Event Type
Malfunction
Date Received
April 28, 2023
Date of Event
June 27, 2019
Report Date
July 21, 2023
Manufacturer
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
Product Code
PBZ
PMA / PMN Number
K163507
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: FOR BACKGROUND, STRYKER ACQUIRED GAUSS SURGICAL INC. (¿GAUSS¿) ON SEPTEMBER 1, 2021. THE MAJORITY OF THE LATE MDRS RESULTED FROM A RETROSPECTIVE REVIEW OF REPORTABILITY DECISIONS MADE PRIOR TO THE ACQUISITION. ADDITIONALLY, STRYKER HAS DETERMINED THAT A CAPA IS NEEDED TO IMPROVE THE POST-ACQUISITION COMPLAINT REVIEW PROCESS ¿ AS A RESULT, CAPA #3338129 WAS APPROVED ON JUNE 14, 2023 AND OPENED ON JUNE 19.

Description of Event or Problem · 0

PER THE CUSTOMER, THEY ARE GETTING READINGS FROM THE SUCTION CANISTERS IN THE OR OF 10CC QBL AND A COUPLE MORE OF <100CC OF QBL. ONE OF THOSE HAD A DROP IN HEMOGLOBIN OF 2 PT. PER THE CUSTOMER THE ACCURACY DOES NOT SEEM ACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

Description of Event or Problem · 0

PER THE CUSTOMER, THEY ARE GETTING READINGS FROM THE SUCTION CANISTERS IN THE OR OF 10CC QBL AND A COUPLE MORE OF <100CC OF QBL. ONE OF THOSE HAD A DROP IN HEMOGLOBIN OF 2 PT. PER THE CUSTOMER THE ACCURACY DOES NOT SEEM ACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1335736 TRITON CANISTER IMAGE PROCESSING DEVICE FOR ESTIMATION OF EXTERNAL BLOOD LOSS PBZ STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP

Patients

Seq Age Sex Outcome Treatment
1 Unknown