PORTEX PNEUPAC VENTILATOR SUPPORT GAUGE
Report
- Report Number
- 3012307300-2023-04686
- Event Type
- Malfunction
- Date Received
- April 27, 2023
- Report Date
- May 10, 2024
- Manufacturer
- SMITHS MEDICAL ASD, INC.
- Product Code
- BXR
- UDI-DI
- 10788942560604
- PMA / PMN Number
- K873498
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A SAMPLE WAS RECEIVED FOR EVALUATION. THE SAMPLE WAS VISUALLY EVALUATED, AND EVENT WAS CONFIRMED THAT THE BLACK INDICATOR NEEDLE ON THE GAUGE READS -7.5 CMH20. PER OUR MANUFACTURING PROCEDURES THE OPERATOR IS SUPPOSED TO 100% VISUALLY INSPECT THE GAUGE READS ZERO WITH 100%-INSPECT CONFIRMATION OF FUNCTIONALITY. THE INSTRUCTIONS FOR USE FOR THE PRODUCT STATE THE USER IS ALLOWED TO TAKE THE FACEPLATE OFF AND MOVE THE GAUGE TO "0" (ZERO) WITH A SCREWDRIVER. THE RETURNED PRODUCT WAS TESTED FOR FUNCTIONALITY AND ACCURACY USING A PRESSURE AND VACUUM TEST. THE FACEPLATE WAS REMOVED, THE GAUGE MOVED TO 0, AND THEN TESTED. THE PRODUCT MET SPECIFICATIONS; THEREFORE, THIS DEFECT CANNOT BE CONFIRMED. THE ROOT CAUSE IS ASSEMBLY ERROR. THE PRODUCT'S HISTORY RECORDS WERE REVIEWED AND THERE WERE NO NON-CONFORMANCES THAT WOULD HAVE RESULTED IN THE REPORTED COMPLAINT. FOR ALL ENQUIRIES OR FOLLOW-UP QUESTIONS RELATED TO THE RECORD, DO NOT USE (B)(6). LOCATED IN SECTIONS G.1., PLEASE DIRECT THOSE TO THE FOLLOWING: (B)(6).
OTHER, OTHER TEXT: B3: DATE OF EVENT, D4: LOT NUMBER, EXPIRATION DATE, AND H4: MANUFACTURE DATE ARE UNKNOWN, NO INFORMATION HAS BEEN PROVIDED TO DATE. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN REQUIRED.
IT WAS REPORTED THAT THE GAUGE DOES NOT READ CORRECTLY AND IS NOT SET FOR ZERO (0). PATIENT INVOLVEMENT IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1045621 | PORTEX PNEUPAC VENTILATOR SUPPORT GAUGE | METER, AIRWAY, PRESSURE (INSPIRATORY FORCE) | BXR | SMITHS MEDICAL ASD, INC. | 55-6060 | 10788942560604 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |