PAXGENE® BLOOD RNA TUBE
Report
- Report Number
- 9617032-2023-00493
- Event Type
- Malfunction
- Date Received
- April 26, 2023
- Date of Event
- April 4, 2023
- Report Date
- April 6, 2023
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- NTW
- UDI-DI
- 54053727621653
- PMA / PMN Number
- K082150
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR ADDITIVE ABNORMALITY WAS OBSERVED. ADDITIONALLY, 100 RETENTION SAMPLES FROM BD INVENTORY WERE VISUALLY INSPECTED AND NO ISSUES WERE OBSERVED; ALL WERE WITHIN SPECIFICATION. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR AN ADDITIVE ABNORMALITY. AN EXACT ROOT CAUSE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION PERFORMED AND SINCE RETENTION SAMPLES AND DHR WERE ALL WITHIN SPECIFICATION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT AFTER ADDITION OF QSX+PK MIX TO PAXGENE® BLOOD RNA TUBE THE ADDITITVE TURNS A DARK GREY. PATIENT IMPACT HAS NOT YET BEEN REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT LIQUID TURNS DARK GREY AND MORE VISCOUS AFTER ADDITION OF QSX+PK MIX..
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1206394 | PAXGENE® BLOOD RNA TUBE | RNA PREANALYTICAL SYSTEMS | NTW | BECTON, DICKINSON AND COMPANY (BD) | 762165 | 0204817 | 54053727621653 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |