BD E-Z SCRUB¿
Report
- Report Number
- 1710034-2023-00447
- Event Type
- Malfunction
- Date Received
- April 26, 2023
- Date of Event
- April 5, 2023
- Report Date
- July 12, 2023
- Manufacturer
- CAREFUSION, INC
- Product Code
- GEC
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ONE SAMPLE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION OF THE SAMPLE IDENTIFIED FOREIGN MATTER ON THE PRODUCT WHICH VERIFIES THE REPORTED ISSUE. FTIR TESTING WAS DONE AT THE VERNON HILLS LOCATION TO IDENTIFY THE FOREIGN MATERIAL. THE TEST RESULTS STATE THAT THE MATERIAL IS A TYPE OF POLYPROPYLENE. LOW DENSITY POLYPROPYLENE IS A MATERIAL USED TO MAKE THE SCRUB BRUSH. OCCASIONALLY, A LOOSE PARTICLE FROM THE MOLDING PROCESS REMAINS WITH THE BRUSH THROUGHOUT THE PRODUCTION AND PACKAGING PROCESS WHICH IS THE MOST PROBABLE SOURCE OF THIS ISSUE. BATCH RECORD REVIEW SHOWS THAT ALL ATTRIBUTE DATE PASSED QUALITY CHECKS THROUGHOUT THE PRODUCTION OF THE LOT, AND NO DEVIATIONS OR DISCREPANCIES WERE NOTED THAT WOULD CONTRIBUTE OR CAUSE THIS ISSUE. REVIEW OF RISK MANAGEMENT DOCUMENT SHOWS THAT SEVERITY RATING IS MODERATE FOR THIS DEFECT WITH POTENTIAL EFFECTS OF FAILURE BEING CUSTOMER DISSATISFACTION AND PATIENT INFECTIONS/COMPLICATIONS. NO FURTHER ACTIONS ARE REQUIRED. COMPLAINTS WILL CONTINUE TO BE TRACKED AND TRENDED.
ON 26APR202: PR (B)(4), INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. IMDRF ANNEX E CODE HEALTH EFFECT CLINICAL CODE: E2403. IMDRF ANNEX A CODE MEDICAL DEVICE PROBLEM CODE: A18.
IT WAS REPORTED BY THE CUSTOMER THEY FOUND MOLD OR SOMETHING GROWING INSIDE, WHEN THE PRODUCT WAS OPENED. VERBATIM: BELOW IS INFO & PHOTOS OF A SCRUB PACK WITH MOLD OR SOMETHING GROWING INSIDE WHEN IT WAS OPENED. LOOKS LIKE MOLD IN UPPER LEFT CORNER OF THE PACK. ADDITIONAL INFORMATION RECEIVED 07-APR-2023. WHAT IS THE DATE OF EVENT? 4/5/23. THE LOT NUMBER PROVIDED OF 2305243, WAS NOT ABLE TO BE FOUND WITHIN OUR SYSTEM FOR THE PRODUCT WITH MATERIAL # 371163. UPON REVIEW OF THE ATTACHED PHOTO OF THE PRODUCT, WE DID OBSERVE THE LOT NUMBER AND FOUND IT TO BE 2305248, WHICH WAS ALSO SEARCHED WITHIN OUR SYSTEM BUT WAS FOUND FOR TO BE A MATCH FOR THE MATERIAL # 371163. CAN YOU PLEASE RE CHECK THE PRODUCT PACKAGING AND CONFIRM THE LOT NUMBER PREVIOUSLY PROVIDED VS OUR FINDINGS? LOT # IS 2305248.
IT WAS REPORTED BY THE CUSTOMER THEY FOUND MOLD OR SOMETHING GROWING INSIDE, WHEN THE PRODUCT WAS OPENED. VERBATIM: BELOW IS INFO & PHOTOS OF A SCRUB PACK WITH MOLD OR SOMETHING GROWING INSIDE WHEN IT WAS OPENED. LOOKS LIKE MOLD IN UPPER LEFT CORNER OF THE PACK. ADDITIONAL INFORMATION RECEIVED 07-APR-2023. WHAT IS THE DATE OF EVENT? (B)(6) 2023. THE LOT NUMBER PROVIDED OF 2305243, WAS NOT ABLE TO BE FOUND WITHIN OUR SYSTEM FOR THE PRODUCT WITH MATERIAL # 371163. UPON REVIEW OF THE ATTACHED PHOTO OF THE PRODUCT. WE DID OBSERVE THE LOT NUMBER AND FOUND IT TO BE 2305248, WHICH WAS ALSO SEARCHED WITHIN OUR SYSTEM BUT WAS FOUND FOR TO BE A MATCH FOR THE MATERIAL # 371163. CAN YOU PLEASE RE CHECK THE PRODUCT PACKAGING AND CONFIRM THE LOT NUMBER PREVIOUSLY PROVIDED VS OUR FINDINGS? LOT # IS 2305248.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1163579 | BD E-Z SCRUB¿ | IMPREGNATED PREOPERATIVE SCRUB BRUSHES (3% CHLOROXYLENOL (PCMX) | GEC | CAREFUSION, INC | 2305248 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |