FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 2.0 TEST KIT 24T OUS

MDR report key: 16808923 · Received April 25, 2023

Report

Report Number
1221359-2023-00653
Event Type
Malfunction
Date Received
April 25, 2023
Date of Event
April 19, 2023
Report Date
June 12, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D4: THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 193-000 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. INVESTIGATION CONCLUSION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M222905 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 193-000 / LOT M222905, TEST BASE PART NUMBER 192-430 / LOT M222905. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M222905 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Additional Manufacturer Narrative · 0

D4: THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 193-000 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY FOR MULTIPLE TESTS ON THE SAME PATIENT PERFORMED ON (B)(6) 2023. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF THREE (3). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY ON A DIRECT TESTED NASAL SAMPLE COLLECTED USING A KITTED SWAB. CONFIRMATION PCR TESTING (UNKNOWN PLATFORM) WAS PERFORMED ON AN UNKNOWN SAMPLE TYPE AND GENERATED NEGATIVE RESULTS. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 2.0 ASSAY FOR MULTIPLE TESTS ON THE SAME PATIENT PERFORMED ON (B)(6) 2023. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF THREE (3). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY ON A DIRECT TESTED NASAL SAMPLE COLLECTED USING A KITTED SWAB. CONFIRMATION PCR TESTING (UNKNOWN PLATFORM) WAS PERFORMED ON AN UNKNOWN SAMPLE TYPE AND GENERATED NEGATIVE RESULTS. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1210339 ID NOW COVID-19 2.0 TEST KIT 24T OUS REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M222905

Patients

Seq Age Sex Outcome Treatment
1 Unknown