FDA Adverse Event Injury Summary report: N

BLUNT TIP SCREW, 4X40MM

MDR report key: 16805461 · Received April 25, 2023

Report

Report Number
0009613350-2023-00189
Event Type
Injury
Date Received
April 25, 2023
Date of Event
March 10, 2023
Report Date
May 8, 2023
Manufacturer
ZIMMER SWITZERLAND MANUFACTURING GMBH
Product Code
HSB
UDI-DI
00889024505445
PMA / PMN Number
K200814
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). D10 ¿ MEDICAL DEVICES: BLUNT TIP SCREW, 4X42MM; ITEM# 47-2486-042-40; LOT# 3082110, BLUNT TIP SCREW, 4X42MM; ITEM# 47-2486-042-40; LOT# 3091524, ANN PH NAIL RT 8.5X240MM; ITEM# 47-2496-240-08; LOT# 3091664, TORQUE LIMITING HANDLE; ITEM# 27923; LOT# UNKNOWN. G2 ¿ FOREIGN: JAPAN. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0009613350 - 2023 - 00190, 0009613350 - 2023 - 00191, 0009613350 - 2023 - 00192. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS. DEVICES ARE USED FOR TREATMENT. MEDICAL RECORDS WERE NOT PROVIDED. IT CAN BE ASSUMED THAT MULTIPLE FACTORS, RELATED TO EITHER PATIENT CONDITION, BEHAVIOR OR IMPLANTATION PROCEDURE, LED OR CONTRIBUTED TO THE MIGRATION OF THE SCREWS. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE LOCKING MECHANISM OF THE NAIL AS A POTENTIAL CONTRIBUTING FACTOR. AS PART OF THE DEEPER INVESTIGATION, A SCIENTIFIC LITERATURE SEARCH WAS CONDUCTED AND A SYSTEMATIC REVIEW OF THE OUTCOME OF INTRAMEDULLARY NAILING FOR ACUTE PROXIMAL HUMERUS FRACTURE SHOWED THAT SCREW MIGRATION AND PERFORATION INTO THE JOINT IS THE SECOND MOST COMMON COMPLICATION FOLLOWING SECONDARY LOSS OF REDUCTION. IN ADDITION, SCIENTIFIC LITERATURE OF COMPETITOR AND PREDICATE DEVICES WERE ASSESSED. THIS REVIEW HAS SHOWN THAT THE PROXIMAL HUMERAL SYSTEM PERFORMS BETTER AND/OR IN EQUAL RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT UNDERWENT INITIAL SURGERY AND APPROXIMATELY AFTER 6 WEEKS THE SURGEON FOUND THAT THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND A REVISION SURGERY WAS PERFORMED TO REMOVE THESE SCREWS. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
489474 BLUNT TIP SCREW, 4X40MM ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES HSB ZIMMER SWITZERLAND MANUFACTURING GMBH N/A 3091511 00889024505445

Patients

Seq Age Sex Outcome Treatment
1 Unknown Hospitalization| R