FDA Adverse Event Malfunction Summary report: N

LUXSTAR CEILING-MOUNT LIGHT

MDR report key: 1679461 · Received May 6, 2010

Report

Report Number
3004517290-2010-00001
Event Type
Malfunction
Date Received
May 6, 2010
Date of Event
April 7, 2010
Report Date
April 7, 2010
Manufacturer
DENTAL EQUIPMENT, LLC (DBA NEWBERG MANUFACTURING)
Product Code
EAZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

RESULTS: THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAD ITS 3 METAL TABS BENT WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. AS A RESULT, THE LENS HEAT SHIELD/COVER WOULD NOT BE ABLE TO BE RE-INSTALLED PROPERLY DURING ROUTINE MAINTENANCE OF THE DENTAL LIGHT. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB. THE DR INFORMED US THAT HE HAD BEEN INFORMED TO BEND THESE TABS IN THE PAST BY HIS INDEPENDENT SERVICE PROVIDER DURING HIS ROUTINE MAINTENANCE OF THE LIGHT ASSEMBLY. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE.

Description of Event or Problem · 1

A DOCTOR WAS PERFORMING A ROUTINE DENTAL PROCEDURE WHEN THE LENS HEAT SHIELD/HOLDER FELL OFF THE LIGHT AND HIT THE PT ON THE FACIAL AREA CAUSING THE PT'S LIP TO SWELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LUXSTAR CEILING-MOUNT LIGHT DENTAL OPERATING LIGHT EAZ DENTAL EQUIPMENT, LLC (DBA NEWBERG MANUFACTURING) CL1000