FDA Adverse Event Death Summary report: N

COGNIS

MDR report key: 1677983 · Received May 6, 2010

Report

Report Number
2124215-2010-10317
Event Type
Death
Date Received
May 6, 2010
Date of Event
April 9, 2010
Report Date
April 9, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUR REVIEW OF THE STORED DATA DOWNLOADED FROM THE DEVICE SHOWED THERE WERE 28 EPISODES REGISTERED OVER A 57-MINS PERIOD APPROX SIX HOURS BEFORE THE REPORTED TIME OF DEATH, WITH A TOTAL OF FIVE HIGH-VOLTAGE SHOCKS AND SEVEN ANTI-TACHYCARDIA PACING BURSTS DELIVERED IN THAT PERIOD. OVERALL, THERE WERE SHORT PERIODS OF UNDERSENSING SEEN WITH HIGH-VOLTAGE SHOCKS DELIVERED INTO WHAT APPEARED TO BE INCESSANT VENTRICULAR FIBRILLATION (VF). THERE WERE ALSO LONG PERIODS WHERE THE RIGHT VENTRICULAR AND SHOCK ELECTROGRAMS SEEMED TO OCCUR AT DIFFERING INTERVALS, PROBABLY RELATED TO ENTRANCE BLOCK OR PARASYSTOLIC ACTIVITY. THE DELIVERED SHOCKS CONVERTED THE VF INTO ASYSTOLE AND AV (ATRIO-VENTRICULAR) PACED BEATS. THESE CONVERSIONS WERE OF SHORT DURATION, HOWEVER, THE VF APPEARED TO RETURN AFTER ONLY A FEW SECONDS. THERAPY DELIVERY WAS COMPLICATED BY THE DEVICE'S VT-1 ZONE PROGRAMMING (MONITOR ONLY) AND THE INTERMITTENT VENTRICULAR SENSING FOR THE ABOVE-NOTED ISSUES. OUR POST MARKET QUALITY ASSURANCE LABORATORY ANALYSIS OF THE RETURNED DEVICE IS PENDING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFO FROM A BOSTON SCIENTIFIC FIELD REP THAT THIS CARDIAC RESYNCHRONIZATION THERAPY-DEFIBRILLATOR (CRT-D) WAS ASSOCIATED WITH A REPORT OF THE PT'S DEATH, AND SUBSEQUENT DEVICE INTERROGATION SHOWED POSSIBLE VENTRICULAR UNDERSENSING. NO ADDITIONAL INFO WAS IMMEDIATELY AVAILABLE FROM THE PT'S PHYSICIAN OR FUNERAL HOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 54 YR Death 4135| 4555| 0184| 4136