MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MMOL/L
Report
- Report Number
- 2032227-2023-193470
- Event Type
- Malfunction
- Date Received
- April 20, 2023
- Date of Event
- April 4, 2023
- Report Date
- June 15, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000521547
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- 003
Narratives
S/W VERSION 6.7V. RETAINER RING= BLACK. CASE TYPE =NGP. CUSTOMER RETURN PUMP DUE TO ALLEGED FAILED BATTERY TEST AND IT WAS FOUND ON 04 APRIL 2023. UNIT PASSED THE DISPLACEMENT TEST, ACTIVE CURRENT TEST, SLEEP CURRENT TEST, AND SELF-TEST. NO FAILED BATTERY ALARM NOTED DURING TESTING. NO ALARMS/ALERT/ANOMALIES NOTED DURING TESTING. PUMP WAS DOWNLOADED USING THUMP FOR HISTORY FILES AND TRACES. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. FAILED BATTERY TEST OCCURRED ON (B)(6) 2023 13:47--14:14. THIS WAS EXPECTED. PUMP ERROR 53 (ESF#3923533 LINE#1232 FILE#14) WAS FOUND ON (B)(6)2023 16:19 IN HISTORY FILES AND TRACES. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND EVIDENCE OF MOISTURE DAMAGE ON THE PCB 1 AND FORCE SENSOR. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED BATTERY TUBE, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, CRACKED BELT CLIP RAILS. P-CAP WAS ATTACHED AND LOCKED INTO PLACE PROPERLY. FAIL BATTERY TEST IS NOT CONFIRMED. PUMP ERROR 53 IS CONFIRMED DUE TO BEING FOUND IN HISTORY FILES AND TRACES AND DUE TO HARDWARE ANOMALY. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE NGP 780G INSULIN PUMP WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE NGP INSULIN PUMP, WHICH IS MARKETED IN THE UNITED STATES SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED PUMP DOESN'T WORK DUE TO BATTERY FAILED ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS NOT UNRESOLVED. THE CUSTOMER WILL DISCONTINUE USING THE PUMP AND THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1613300 | MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MMOL/L | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1885 | HG5ZYYBZZ | 000000763000521547 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |