FDA Adverse Event Malfunction Summary report: N

ON-BOARD IMAGER (OBI)

MDR report key: 1677052 · Received May 5, 2010

Report

Report Number
2916710-2009-00087
Event Type
Malfunction
Date Received
May 5, 2010
Date of Event
May 22, 2009
Report Date
June 17, 2009
Manufacturer
VARIAN MEDICAL SYSTEMS
Product Code
IYE
PMA / PMN Number
K040192
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING FURTHER REVIEW OF THE INITIAL CUSTOMER REPORT, IT WAS DISCOVERED THAT THE PROBLEM EXISTS WHEN VARIOUS STEPS IN THE (B)(4) ACQUISITION AND MATCH PROCESS OCCUR - SUCH AS LOADING THE REFERENCE CT DATASET, LOADING THE (B)(4) DATASET AND SAVING THE (B)(4) DATASET - COMPUTER MEMORY IS CONSUMED. NOT ALL OF THIS COMPUTER MEMORY IS FREED UP FOR REUSE WHEN THE PT IS CLOSED. THEREFORE, AS MORE PTS ARE SCANNED USING (B)(4), THERE MAY BE INSUFFICIENT MEMORY FOR OBI APPLICATION TO CONTINUE TO OPERATE. VARIAN HAS PROVIDED A CUSTOMER TECHNICAL BULLETIN TO INFORM USERS OF A SOLUTION WHEN THIS PROBLEM OCCURS. IF MEMORY LOG SUGGESTS THAT ALL OF THE FREE MEMORY MAY POTENTIALLY BE CONSUMED, THE SOLUTIONS IS TO CLOSE THE OBI APPLICATION AFTER EVERY 10 (B)(4) SCANS. RESTARTING THE OBI APPLICATION WILL FREE UP THE AVAILABLE MEMORY AND WILL ALLOW THE OBI TO OPERATE NORMALLY. ALTHOUGH THERE WAS NO SERIOUS INJURY REPORTED IN THIS CASE, VARIAN HAS DETERMINED THAT A MDR IS APPROPRIATE, AS THIS MALFUNCTION, SHOULD IT RECUR, COULD RESULT IN UNINTENDED RADIATION TO THE WRONG ANATOMICAL SITE. AN INTERNAL REVIEW FOLLOWING A SUBSEQUENT REPORT OF THIS ISSUE HAS LED TO THE DECISION TO REPORT THIS OLDER REPORT. NO ADD'L FOLLOW-UP TO THIS MDR IS EXPECTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AFTER PERFORMING A (B)(6) MATCH, USER NOTICED A 37.9CM SHIFT ON THE LNG. KNOWING THIS WAS INCORRECT, CUSTOMER CANCELLED THE SHIFT FOR THAT DAY. NO OTHER PT INFO WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ON-BOARD IMAGER (OBI) ACCELERATOR, LINEAR, MEDICAL IYE VARIAN MEDICAL SYSTEMS

Patients

Seq Age Sex Outcome Treatment
1