FIRST PICC¿ S/L PERIPHERALLY INSERTED CENTRAL CATHETER
Report
- Report Number
- 0001625425-2023-01030
- Event Type
- Malfunction
- Date Received
- April 19, 2023
- Date of Event
- March 24, 2023
- Report Date
- May 30, 2023
- Manufacturer
- ARGON MEDICAL DEVICES
- Product Code
- FOZ
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN RECEIVED AND REVIEWED.
A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE OPENED CATHETER WAS RETURNED FOR REVIEW. VISUAL INSPECTION OF THE SAMPLE CONFIRMED THE SILICONE EXTENSION WAS DETACHED FROM THE CATHETER TUBING JUST BELOW THE INVERTED TRIANGLE BENEATH THE SECUREMENT DISC. THE BREAKAGE SITE EXHIBITED JAGGED EDGES WHICH IS INDICATIVE OF A TENSILE FORCE APPLIED TO THE CATHETER. A DEFINITE ROOT CAUSE FOR THE BREAKAGE IS NOT CERTAIN. DUE TO THE JAGGED EDGES, POSSIBLY AN EXCESSIVE FORCE WAS APPLIED TO THE CATHETER. SINCE THE REPORTED ISSUE WAS MOST LIKELY THE RESULT OF AN EVENT WITHIN THE USER ENVIRONMENT, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. ARGON WILL CONTINUE TO MONITOR FOR ISSUES OF THIS NATURE IN THE FUTURE.
DESCRIBE THE EVENT OR PROBLEM: PATIENT MOVED ARM UP AND ARGON NEONATAL PICC LINE SNAPPED IN HALF CAUSING PATIENT TO LOSE CENTRAL ACCESS. NEONATAL PICC WAS SECURED PER NEONATAL INTENSIVE CARE UNIT POLICY, EXTENSION TUBING BEYOND DRESSING IS WHERE THE LINE SNAPPED IN HALF. PATIENT HAD TOTAL PARENTERAL NUTRITION (TPN), LIPIDS, AND PROSTAGLANDINS (PGE) RUNNING THROUGH LINE. NEONATAL PICC PULLED FROM PATIENT'S ARM, LINE INTACT AND SAVED IN BIOHAZARD BAG. PATIENT GOING FOR OPEN HEART IN OR THIS MORNING. PATIENT ON PEDIALYTE FOR FEEDS AFTER MIDNIGHT AND NOTHING BY MOUTH (NPO). PATIENT REQUIRED IV 10% DEXTROSE SOLUTION (D10) TO REPLACE TPN, AND NOW REQUIRING Q3 GLUCOSE CHECKS.
DESCRIBE THE EVENT OR PROBLEM: PATIENT MOVED ARM UP AND ARGON NEONATAL PICC LINE SNAPPED IN HALF CAUSING PATIENT TO LOSE CENTRAL ACCESS. NEONATAL PICC WAS SECURED PER NEONATAL INTENSIVE CARE UNIT POLICY, EXTENSION TUBING BEYOND DRESSING IS WHERE THE LINE SNAPPED IN HALF. PATIENT HAD TOTAL PARENTERAL NUTRITION (TPN), LIPIDS, AND PROSTAGLANDINS (PGE) RUNNING THROUGH LINE. NEONATAL PICC PULLED FROM PATIENT'S ARM, LINE INTACT AND SAVED IN BIOHAZARD BAG. PATIENT GOING FOR OPEN HEART IN OR THIS MORNING. PATIENT ON PEDIALYTE FOR FEEDS AFTER MIDNIGHT AND NOTHING BY MOUTH (NPO). PATIENT REQUIRED IV 10% DEXTROSE SOLUTION (D10) TO REPLACE TPN, AND NOW REQUIRING Q3 GLUCOSE CHECKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1937216 | FIRST PICC¿ S/L PERIPHERALLY INSERTED CENTRAL CATHETER | FIRST PICC | FOZ | ARGON MEDICAL DEVICES | UNK | 11452818 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other| R |