FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 8MM

MDR report key: 16768735 · Received April 19, 2023

Report

Report Number
0002023141-2023-01118
Event Type
Injury
Date Received
April 19, 2023
Date of Event
January 9, 2023
Report Date
September 26, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019546
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). G4: PMA/510(K) NUMBER : K011028 AND K013227.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). H3: DEVICE EVALUATED BY MANUFACTURER: CHANGE ¿NO' TO 'YES.' A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1223879. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1223879 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿BONE LOSS.¿

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT WENT TO THE CLINIC WITH IMPLANT AND CROWN COMPLETELY REMOVED DUE TO BONE LOSS. PROCEDURE WAS NOT COMPLETED WITH OTHER IMPLANT

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1373620 IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 8MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB8 1223879 00889024019546

Patients

Seq Age Sex Outcome Treatment
1 62 YR Prefer Not To Disclose Required Intervention