FDA Adverse Event Malfunction Summary report: N

ENDURANT STENT GRAFT

MDR report key: 16768512 · Received April 19, 2023

Report

Report Number
9612164-2023-01668
Event Type
Malfunction
Date Received
April 19, 2023
Date of Event
October 25, 2022
Report Date
April 19, 2023
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC RECEIVED THE FOLLOWING INFORMATION FROM A JOURNAL ARTICLE ENTITLED; EVOLUTION OF OPEN ANEURYSMORRHAPHY FOR MANAGEMENT OF SAC EXPANSION AFTER ENDOVASCULAR REPAIR OF ABDOMINAL AORTIC ANEURYSMS OZAWA H., OHKI T., SHUKUZAWA K., CHONO Y., OMORI M., BABA T., HARA M., TACHIHARA H. J. VASC. SURG.. 2023 MAR 01; 77(3):760-768. DOI: 10.1016/J.JVS.2022.10.024 MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ENDURANT AND NON MDT STENT GRAFTS WERE IMPLANTED IN EVAR PROCEDURES ON UNKNOWN DATES. FOLLOWING THIS, 28 PATIENTS UNDERWENT OPEN AN EURYSMORRHAPHY (OA) FOR THE TREATMENT OF NON RUPTURED SAC EXPANSION OVER A SIX YEAR PERIOD. OPEN ANEURYSMORRHAPHY INVOLVES SAC EXPOSURE WITHOUT DISSECTION OF THE PROXIMAL OR DISTAL NECK, SACOTOMY AND LIGATION OF BACKBLEEDING VESSELS, PRESERVATION OF THE PRIOR STENT GRAFT, AND TIGHT CLOSURE OF THE SAC AROUND THE STENT GRAFT. THE MEDIAN SAC DIAMETER AT THE OPEN ANEURYSMORRHAPHY WAS 79 MM. THE MEDIAN DURATION FROM THE INDEX EVAR TO THE OA WAS 82 MONTHS. THE FOLLOWING MALFUNCTIONS WERE IDENTIFIED IN PATIENTS FOLLOWING THE EVAR PROCEDURES: TYPE IA, IB, IIIB , IV ENDOLEAKS, INADEQUATE SEAL , MIGRATION THE FOLLOWING ADVERSE EVENTS WERE REPORTED FOLLOWING THE OA PROCEDURES; ABDOMINAL LYMPHORRHEA (RESOLVED WITH CONSERVATIVE TREATMENT) , FAILED HEMOSTASIS OF THE COMMON FEMORAL ARTERY REQUIRING SURGICAL REPAIR, BLOOD LOSS REQUIRING A BLOOD TRANSFUSION, ANEURYSM ENLARGEMENT, INFECTION, FISTULA PATIENT MORTALITY WAS REPORTED BUT THERE IS NO CAUSAL LINK THAT A ENDURANT STENT GRAFT CAUSED OR CONTRIBUTED TO ANY DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1443764 ENDURANT STENT GRAFT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR MIH MEDTRONIC IRELAND UNK-CV-SR-ENDURANT

Patients

Seq Age Sex Outcome Treatment
1 Unknown