FDA Adverse Event Malfunction Summary report: N

CITADEL PATIENT CARE SYSTEM

MDR report key: 16768206 · Received April 19, 2023

Report

Report Number
3007420694-2023-00082
Event Type
Malfunction
Date Received
April 19, 2023
Date of Event
March 21, 2023
Report Date
June 15, 2023
Manufacturer
ARJOHUNTLEIGH POLSKA SP. Z O.O.
Product Code
IOQ
UDI-DI
05055982755764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE AVAILABLE UPON INVESTIGATION CONCLUSION.

Additional Manufacturer Narrative · 0

FOLLOWING THE INFORMATION PROVIDED THE PATIENT WITH PREEXISTING STAGE 4 PRESSURE INJURY ON THE SACRUM WAS LAYING ON THE DEFLATED MATTRESS. THE CUSTOMER REPORTED THAT THE PREEXISTING PRESSURE INJURY HAS GOTTEN WORSE OVERNIGHT. THE DEVICE WAS EVALUATED AFTER THE EVENT. NO DEVICE MALFUNCTION WAS FOUND. DURING THE INTERVIEW WITH CUSTOMER, IT WAS CONFIRMED THAT THE CAREGIVERS DID NOT KNOW HOW TO USE THE BED CONTROLS FUNCTION WHICH CAUSED THE MATTRESS DEFLATION. THE INSTRUCTION FOR USE (FU) FOR CITADEL PATIENT THERAPY SYSTEM PROVIDES A EXPLANATION ON HOW TO OPERATE THE DEVICE, INCLUDING A STEP-BY-STEP GUIDE ON ATTACHING THE AIR MATTRESS. ADDITIONALLY, THERE IS A DEDICATED SECTION WITHIN THE IFU THAT PROVIDES AN EXPLANATION OF THE FUNCTIONS OF EACH MATTRESS CONTROL PANEL BUTTON. ADDITIONALLY, THE DOCUMENT PROVIDES TROUBLESHOOTING INFORMATION TO ASSIST CAREGIVERS IN RESOLVING THE PROBLEM RELATED TO MATTRESS DEFLATION. PRESSURE INJURIES ARE COMPLEX AND ARE A RESULT OF MANY FACTORS INCLUDING: ADVANCED AGE, IMMOBILITY, CO-MORBIDITIES, MICROCLIMATE, INCONTINENCE AND LACK OF BEING TURNED OR REPOSITIONED FREQUENTLY ENOUGH. THE IFU FOR CITADEL BED (830.238-EN) INCLUDES THE FOLLOWING INFORMATION RELATED TO THE SUBJECT OF THE INVESTIGATION: - "MONITOR SKIN CONDITIONS REGULARLY AND CONSIDER ADJUNCT OR ALTERNATIVE THERAPIES FOR HIGH ACUITY PATIENTS. GIVE EXTRA ATTENTION TO SKIN OVER ANY RAISED SIDE BOLSTER AND TO ANY OTHER POSSIBLE PRESSURE POINTS AND LOCATIONS WHERE MOISTURE OR INCONTINENCE MAY OCCUR OR COLLECT. EARLY INTERVENTION MAY BE ESSENTIAL TO PREVENTING SKIN BREAKDOWN." THE DEVICE WAS USED FOR A PATIENT TREATMENT WHEN WORSENING OF PREEXISTING INJURY WAS NOTED BY THE CAREGIVERS, AND FROM THAT PERSPECTIVE IT PLAYED A ROLE IN THE EVENT. THERE WAS NO DEVICE MALFUNCTION FOUND WHICH MIGHT LED TO REPORTED MATTRESS DEFLATION. THE COMPLAINT WAS REPORTED IN ABUNDANCE OF CAUTION DUE TO FACT THAT THE PATIENT HAD PREEXISTING STAGE IV PRESSURE INJURY WHICH HAS BEEN WORSENED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE PROVIDED UPON INVESTIGATION CONCLUSION.

Description of Event or Problem · 0

FOLLOWING THE INFORMATION PROVIDED THE PATIENT WAS LAYING ON THE DEFLATED MATTRESS. IT WAS REPORTED THAT THE PATIENT HAD A STAGE 4 PRESSURE INJURY ON HIS SACRUM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1373590 CITADEL PATIENT CARE SYSTEM BED, FLOTATION THERAPY, POWERED IOQ ARJOHUNTLEIGH POLSKA SP. Z O.O. CX811A3F3AMAB0 05055982755764

Patients

Seq Age Sex Outcome Treatment
1 Male Other